Mapping and Outcome Implications of Left Atrial Low Voltage Regions as Determined by a Novel 122-electrode Spherical Mapping and Ablation Catheter.
Atul Verma, Hans Kottkamp, Hugh Calkins, Devi G Nair, Philipp Sommer, Ayman A Hussein, Burghard Schumacher, Srinivas Dukkipati,
McGill University Health Centre, Montreal, Canada. · Westpfalz Hospital, Kaiserslautern, Germany.
Key points
From the paper's conclusion- Conclusions A voltage threshold of 1.0 mV is proposed as a threshold for defining LVA using the spherical catheter, but a range of approximately 0.5-1.5 mV yielded similar performance.
Abstract
AIMS: A novel 122-electrode spherical mapping/pulsed field ablation (PFA) device is known to effectively ablate atrial fibrillation (AF). The aim of this study is threshold optimization of LVA for the unique geometry of the Globe electrodes and the prognostic association, to evaluate if the extent of low voltage areas could predict recurrence. METHODS AND RESULTS: The PULSAR study was an IDE trial of 183 patients undergoing first-time paroxysmal AF ablation with one-year outcomes of any atrial arrhythmia. During this study, a subset of patients (n=82) had detailed pre-ablation voltage maps performed with the catheter. Pulmonary vein isolation (PVI) was then performed with PFA. Patients were followed with weekly transtelephonic monitoring, ECG, and Holters at 3, 6, and 12 months. Recurrence was defined as any atrial arrhythmia (AA) > 30 seconds.Receiver operator curve (ROC) analysis revealed that a low voltage threshold of 1.0 mV had the highest value (0.735). The percentage LVA surface area resulting from a threshold of 1.0 mV that predicted AA recurrence was 11.6%. Univariate and multivariate analysis revealed the percentage LVA was the strongest predictor of AA recurrence. The 12-month freedom from AA was 84.6% and 52.9% for subjects under and over the low voltage area threshold respectively. CONCLUSIONS: A voltage threshold of 1.0 mV is proposed as a threshold for defining LVA using the spherical catheter, but a range of approximately 0.5-1.5 mV yielded similar performance. Increased low voltage area in the pre-ablation atrial map was associated with a higher risk of AA recurrence.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Distribution of left atrial low-voltage areas and its association with recurrent types of atrial tachyarrhythmiaFrontiers in Cardiovascular Medicine · 2026
- Pulsed-Field Ablation to Enhance the Durability of Pulmonary Vein Isolation and Treat Paroxysmal Atrial FibrillationJournal of the American College of Cardiology · 2026
- Optimal near-field unipolar voltage thresholds using a lattice-tip mapping catheter corresponding to bipolar low-voltage areas in atrial fibrillation ablationEuropean Heart Journal · 2026
- Predictive Value of Left Atrial Strain Combined with Low-Voltage Substrate for Recurrence After Radiofrequency Catheter Ablation in Persistent Atrial Fibrillation.Diagnostics (Basel, Switzerland) · 2026
- Left atrial low-voltage areas in persistent atrial fibrillation: from atrial cardiomyopathy marker to selective ablation targetFrontiers in Cardiovascular Medicine · 2026
- Identification of the mapping-based voltage-derived parameter for accurate arrhythmia outcome prediction in patients undergoing a first PVIEuropean Heart Journal · 2026