Optimized workflow for pentaspline PFA and its impact on PVI durability: evidence from systematic remapping
Daniel Rodriguez-Munoz, Lucía Cobarro Gálvez, Ez Alddin Rajjoub Al-Mahdi, Isabel López-Alacid, Javier Ramos Jiménez, A Marco Del Castillo, Luis Borrego Bernabé, Martín Negreira‐Caamaño,
Research Institute Hospital 12 de Octubre · Universidad Complutense de Madrid
Key points
From the paper's conclusion- We conducted a single-centre, prospective study including consecutive patients undergoing first-time P-PFA pulmonary vein isolation (PVI), subsequently scheduled for systematic invasive remapping ≥30 days after the index procedure.
Abstract
AIMS: Real-world series report substantial rates of pulmonary vein (PV) reconnection following index procedure with a pentaspline pulsed-field ablation (P-PFA) catheter. The contrast with initial durability data suggests this may be due to limitations in procedural workflow. METHODS AND RESULTS: We conducted a single-centre, prospective study including consecutive patients undergoing first-time P-PFA pulmonary vein isolation (PVI), subsequently scheduled for systematic invasive remapping ≥30 days after the index procedure. Workflow was progressively refined in three phases, aiming at ensuring adequate tissue contact and based on previous remapping findings. The primary endpoint was PVI durability, per vein and per patient. Of 118 enrolled patients, 94 (Phase 1:n = 30; Phase 2:n = 34; Phase 3:n = 30) underwent systematic remapping, where 13, 10, and 1 reconnected PVs were observed in Phases 1, 2, and 3, respectively. This resulted in per-vein durability of 89.8%, 92.9%, and 99.2% (P = 0.005) and per-patient durability of 60.0%, 76.5%, and 96.7% (P = 0.004). Mean total PFA applications increased from 36.8 ± 5.2 in Phase 1 to 52.1 ± 8.0 and 58.7 ± 15.1 in Phases 2 and 3, respectively (P < 0.001). Median procedure time rose from 43 (39-50) minutes in Phase 1 to 46 (42-56) minutes in Phase 2 and 69 (61-77) minutes in Phase 3 (P < 0.001), with significant parallel increases in LA dwell time (P < 0.001) across phases. No major or minor complications occurred during index or remapping procedures. CONCLUSION: Progressive workflow refinements for pentaspline P-PFA markedly improved PVI durability to nearly 100%, supporting the concept that PV reconnection is predominantly technique-dependent rather than intrinsic to catheter or waveform design.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reversible posterior wall electroporation following pentaspline versus balloon-in-basket pulsed-field ablationEuropean Heart Journal · 2026
- Lesion Durability of the Second Generation Pentaspline Pulsed Field Ablation Catheter NAVIGATE-PF Phase 2 ResultsCirculation Arrhythmia and Electrophysiology · 2026
- Characteristics of pulmonary vein reconnection sites across three pulsed-field ablation systems: findings from repeat proceduresEuropean Heart Journal · 2026
- Early clinical experience with single-catheter map-and-ablate pentaspline PFA workflows in Europe: findings from the OPALISE Sub-Study of the FARAVERSE RegistryEuropean Heart Journal · 2026
- Pulsed-field ablation demonstrates superior pulmonary vein isolation durability at 90 days: Pooled analysis of protocol-mandated remapping dataHeart Rhythm · 2025
- Pulsed electric field for pulmonary vein isolation in paroxysmal atrial fibrillation: A multicenter experience comparing pentaspline and lattice-tip catheters.Heart rhythm O2 · 2026