Pacemaker dependency rates decrease within the first year of post-TAVR implantation
G M Wade, R Patel, A Downs, J Nazari, M Ricciardi, J Levisay, J Rosenberg, A Ro,
Key points
From the paper's conclusion- These findings can help guide pacemaker choices and programming.
Abstract
Introduction High degree AV block occurs in a proportion of patients after TAVR leading to pacemaker implantation. The rate and clinical predictors of pacemaker dependency are not well established. Objectives The objective of this study was to determine the percentage of patients who remain pacemaker dependent at 6 weeks, 6 months, and 12 months after TAVR implantation. Secondary endpoints included patient and procedural predictors of long-term dependency. Methods Consecutive patients from May 2024 to October 2025 who underwent pacemaker placement at a single center within 30 days before or after TAVR were retrospectively included. Pacemaker dependency was defined as >90% ventricular pacing burden on device interrogation at follow-up. Patient comorbidity and procedural data were used to analyze predictors of long-term dependency by Chi-Square testing and Pearson correlation. Results Of 430 patients undergoing TAVR, 37 patients (8.6%) received pacemakers within 30 days. 45.9% of those patients (17/37) remained pacemaker dependent at 6 weeks. Dependency rates decreased to 29.7% (11/37 patients) at 6 months. At 12 months only 16.2% of patients (6/37) remained pacemaker dependent. Pre-procedural conduction abnormalities (left bundle branch block and bifascicular block) and atrial arrhythmias (atrial fibrillation and flutter) were significant predictors of pacemaker dependency. Patients with LBBB pre-TAVR who received pacemakers post-TAVR had a 30% rate of dependency compared to 11% for patients without LBBB (p<0.01). Similar rates were observed in 33% of patients with pre-existing bifascicular block compared to 13% without (p<0.01). Patients with atrial fibrillation and/or flutter had a 25% dependency rate compared to 12% without (p<0.01). Other factors (RBBB, sex, CAD, and LVEF) had no statistically significant association with dependency. The Self-expanding supra-annular valves (n=6) resulted in the highest rate of dependency at 33%, followed by Balloon-expandable intra-annular valves (n=22) at 14%, and Self-expanding intra-annular valves (n=9) at 11%. Analysis of valve size indicates a strong positive relationship (r=0.928) between valve width and 12-month dependency. Conclusion In this single center experience, the rate of periprocedural pacemaker placement was modest at 8.6%. The pacemaker dependency rate was high at 6 weeks (45.9%), but decreased to only 16.2% at 12 months, predicted by pre-existing conduction abnormalities, atrial arrhythmias, and valve type/width. These findings can help guide pacemaker choices and programming.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of TAVR on pacemaker dependency in patients with pre-existing devices: predictors and one-year outcomesEuropean Heart Journal · 2026
- Long-term changes in pacemaker stimulation burden and its influence on clinical outcomes after transcatheter aortic valve implantation: a multicenter studyEuropean Heart Journal · 2026
- Transcatheter aortic valve implantation and need for post-discharge pacemaker implantation in the modern era of fast dischargeEuropean Heart Journal · 2026
- Conduction disturbances, permanent pacing, and long-term outcomes in older adults after transcatheter aortic valve implantation: a retrospective cohort study in ChinaFrontiers in Medicine · 2026
- Impact of permanent pacemaker implantation after transcatheter aortic valve implantation: an updated systematic review with meta-analysis of reconstructed time-to-event dataEuropean Heart Journal · 2026
- Left bundle branch block and risk for pacemaker implantation after TAVI: insights from the SwissTAVI Cohort StudyEuropean Heart Journal · 2026