Diagnostic accuracy of ai-enabled wearables for detecting atrial fibrillation recurrence after catheter ablation: a systematic review and meta-analysis
J Giorgi, R Huntermann, P G Batista, M E Molinari, J Peres de Oliveira, R R A Santos Silva, M V Montenegro, L A Lucena,
Key points
From the paper's conclusion- Conclusion In this meta-analysis of four studies, AI- and device-enabled monitoring approaches showed good overall diagnostic performance for detecting atrial fibrillation recurrence after catheter ablation, with particularly high specificity.
Abstract
Introduction Artificial intelligence (AI)–enabled monitoring has emerged as a promising approach for detecting atrial fibrillation (AF) recurrence after catheter ablation, providing continuous or high-frequency rhythm assessment beyond traditional ECG- and Holter-based follow-up. However, the diagnostic performance of AI-driven wearables and handheld devices in this context has not been systematically evaluated. Purpose We performed a systematic review and meta-analysis to quantify the diagnostic accuracy of AI-based and device-enabled monitoring strategies for identifying post-ablation AF recurrence. Methods A systematic review and meta-analysis were conducted in accordance with PRISMA guidelines. Four studies comprising 333 participants were included, evaluating handheld single-lead ECG systems with AI algorithms, deep learning–enabled symptom-triggered recorders, implantable loop recorders, and wearable photoplethysmography-based approaches, using conventional ECG or continuous monitoring comparators. Pooled diagnostic performance was assessed using random-effects models, univariate sensitivity and specificity analyses with 95% confidence intervals (CIs), and a bivariate SROC model were applied. Results The overall sensitivity was 85.9% (95% CI 76.1–92.1) and specificity was 98.8% (95% CI 84.4–99.9). The pooled positive predictive value was 98.1% (95% CI 81.3–99.8), while the negative predictive value was 89.9% (95% CI 81.3–94.8). The summary area under the receiver operating characteristic curve was 0.93 (95% CI 0.84–0.98), indicating high overall discriminative ability. Substantial heterogeneity was observed across studies, reflecting differences in monitoring modality, patient populations, and recurrence definitions. Conclusion In this meta-analysis of four studies, AI- and device-enabled monitoring approaches showed good overall diagnostic performance for detecting atrial fibrillation recurrence after catheter ablation, with particularly high specificity. Although these findings suggest potential utility as adjunctive tools within post-ablation follow-up, the limited evidence base and between-study heterogeneity highlight the need for larger prospective studies to clarify their role in contemporary surveillance pathways.Graphical Abstract
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Monitoring Atrial Fibrillation Using Wearable Digital Technologies: The Emerging Role of SmartwatchesJournal of Clinical Medicine · 2025
- Wearable Smartwatches for Atrial Fibrillation Detection and Burden Estimation After Ablation: Comparison With Continuous MonitoringEP Europace · 2025
- Artificial Intelligence in Wearable ECG and PPG Devices for Arrhythmia Detection: A Narrative Review : Short title: Artificial Intelligence in Wearable Devices for Arrhythmia Detection.Galen medical journal · 2026
- Feasibility and reliability of a smartwatch to detect atrial fibrillation after cardiac surgery: a prospective studyFrontiers in Digital Health · 2026
- AI-assisted ECG screening for atrial fibrillation in older adults: comparison of short-term and one-week monitoringEuropean Heart Journal · 2026
- Accuracy of Smartwatches in the Detection of Atrial Fibrillation: A Systematic Review and Diagnostic Meta-Analysis.JACC. Advances · 2025