Accuracy of artificial intelligence-enabled handheld ultrasound devices to assess left ventricular ejection fraction: A systematic review and meta-analysis

J Winarta, A Gupta, S Zahedmohajerani, R Burgess, C Smith, H Marie North, N Samad, K Witte,

Meta-analysis

Key points

From the paper's conclusion
  • Overall, AI correctly classified examinations most commonly when utilising 30% thresholds (accuracy 0.96 [0.95 – 0.98]); it performed worst in 40% thresholds (accuracy 0.85 [0.76 – 0.94], I2 = 96.19%).

Abstract

Introduction Transthoracic echocardiography (TTE) underpins the diagnosis, phenotyping and subsequent management of heart failure (HF). However, access to high-quality cardiac imaging remains inequitable. Globally, workforce shortages, rising HF prevalence, and increasing service demand contribute to diagnostic delays and variation in care.1-2 Artificial intelligence (AI)-enabled handheld ultrasound devices (HUDs) have emerged as a potential solution, offering the possibility of streamlining or decentralising cardiac evaluation. Despite growing clinical uptake, the diagnostic accuracy of AI-enabled HUDs, particularly when used by novice operators and in real-world clinical settings, has not been comprehensively established. Purpose We conducted a systematic review and meta-analysis to determine the diagnostic accuracy of commercially available AI-enabled HUD compared with TTE in determining reduced left ventricular ejection fraction (LVEF). Methods The protocol was registered with PROSPERO. We searched PubMed, MEDLINE, Embase, and CENTRAL from inception to 1 February 2026 for studies comparing AI-enabled HUDs with standard echocardiography in adult patients. Two reviewers independently screened and extracted data; studies were included only if 2×2 confusion matrices could be derived for AI-enabled HUD versus TTE. Where this was not possible, but the study appeared potentially relevant, we made requests to corresponding authors. Risk of bias and applicability were assessed using QUADAS-2. All statistical analyses were performed using STATA 19 (StataCorp. 2025. Stata Statistical Software: Release 19. College Station, TX: StataCorp LLC.). We pooled diagnostic accuracy using random-effects and hierarchical summary receiver operating characteristic (HSROC), reported 95% confidence intervals and I², and analysed subgroups by operator experience and LVEF cut-off. Results Seventeen studies including 2,879 examinations were included. There was variability across studies in the threshold for reduced LVEF used for diagnostic accuracy analyses. The overall sensitivity was 0.80 (0.74 – 0.87, I2 = 86.34%), specificity 0.93 (0.91 – 0.95, I2 = 82.24%), PPV 0.80 (0.74 – 0.86, I2 = 85.85%), NPV 0.91 (0.87 – 0.95, I2 = 97.02%), and accuracy 0.90 (0.87 – 0.93, I2 = 90.86%). Sensitivity was higher for novices (0.91 [0.85–0.98]) compared to experts (0.70 [0.60–0.81]). Overall, AI correctly classified examinations most commonly when utilising 30% thresholds (accuracy 0.96 [0.95 – 0.98]); it performed worst in 40% thresholds (accuracy 0.85 [0.76 – 0.94], I2 = 96.19%). Conclusion AI-enabled HUD demonstrated excellent performance for identifying reduced LVEF, supporting its use as a screening and triage tool. However, substantial inter-study heterogeneity and operator-dependent differences indicate performance is context-specific and should be interpreted with training, scan protocol, and manufacturer information.HSROC CurvePRISMA Flow Chart

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers