B-409 Assay Development and Validation of Therapeutic Drug Monitoring (TDM) of Busulfan on Both Single Quadrupole and Triple Quadrupole LC-MS in Human Plasma
Yinghua Qiu, Timothy Kee, Rosa Yeh, Riley Rogan, Jenna Davis, Hung Nguyen, Avadhut Joshi, Sung Choi
Key points
From the paper's conclusion- Therefore, this study aims to develop and validate a robust and sensitive HPLC–MS method on both single quadrupole and triple quadrupole platforms for the quantification of busulfan in human plasma.
Abstract
Background Busulfan is a bifunctional alkylating agent that cross-links DNA, disrupting replication and leading to apoptosis. It is a cornerstone of combination conditioning regimens used prior to hematopoietic cell transplantation (HCT) and as a single agent prior to gene therapy. Accurate measurement of busulfan concentrations in plasma is essential for characterizing its pharmacokinetic profile and optimizing individualized dosing. Therefore, this study aims to develop and validate a robust and sensitive HPLC–MS method on both single quadrupole and triple quadrupole platforms for the quantification of busulfan in human plasma. Methods In this study, two robust reverse-phase HPLC methods coupled with single quadrupole and triple quadrupole mass spectrometry were developed and validated for the quantitative determination of busulfan in human plasma. Analytes were extracted from 50 µL of human plasma using protein precipitation with acetonitrile. The processed samples were analyzed on an Agilent Technologies 6470 Triple Quadrupole LC-MS/MS system and an Agilent Technologies Infinity Lab LC/MSD iQ single-quadrupole system, both equipped with a ZORBAX Eclipse Plus C18 RRHD column (3.0 × 50 mm, 1.8 µm) and a 3.0 × 5 mm guard column of the same packing material. For the single quadrupole LC-MS, a 1.6-minute isocratic method was performed at 68% mobile phase A (2 mM ammonium acetate in water containing 0.05% formic acid and 2% acetonitrile) and 32% mobile phase B (acetonitrile) with a flow rate of 0.45 mL/min. For the triple quadrupole LC-MS, a 1.6-minute isocratic method was performed at 50% mobile phase A and 50% mobile phase B at the same flow rate of 0.45 mL/min. The injection volume was 2.0 µL. Results Method validation was performed in accordance with Clinical and Laboratory Standards Institute C62-Ed2, U.S. Food and Drug Administration Bioanalytical Method Validation guidance, European Medicines Agency Bioanalytical Method Validation guidance, and Scientific Working Group for Forensic Toxicology (SWGTOX) Standard Practices for Method Validation in Forensic Toxicology. The analytical measurement range (AMR) for busulfan was established at 50–5,000 ng/mL, with the clinical reportable range (CRR) extending up to 10,000 ng/mL with validated 2-fold and 5-fold dilutions. Both assays demonstrated excellent performance, including system suitability and linearity (R² > 0.999), accuracy (recoveries 94%–102%), and precision (within-run and between-run CV% = 6%). Specificity was confirmed, with interference from matrix components, internal standard, and commonly co-administered drugs = 20% of the LLOQ. Carryover was < 15% of the LLOQ. The methods also showed high sensitivity, with a limit of detection (LOD) of 5 ng/mL and a lower limit of quantification (LLOQ) of 50 ng/mL. Conclusion Robust HPLC methods coupled with single quadrupole and triple quadrupole mass spectrometry were successfully developed and validated for the quantification of busulfan in human plasma. Both methods demonstrated excellent analytical performance, including strong linearity, low imprecision, high accuracy, minimal carryover, negligible matrix effects, and acceptable stability under defined conditions. These methods are suitable for clinical pharmacokinetic studies and real-time therapeutic drug monitoring, enabling targeted optimization of busulfan dosing in both pediatric and adult patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A rapid isocratic LC-MS/MS method for the quantification of busulfan in human plasma and its clinical application in pediatric bone marrow transplantation.Analytical methods : advancing methods and applications · 2026
- B-410 Assay Development and Validation for Therapeutic Drug Monitoring (TDM) of Treosulfan on LC-MS/MS in Human PlasmaClinical Chemistry · 2026
- Population pharmacokinetics and model-informed precision dosing of busulfan in Chinese pediatric allogeneic hematopoietic stem cell transplantation: a prospective real-world cohort study.Frontiers in pharmacology · 2026
- Development and Validation of an HPLC-MS/MS Method for Determining BMS-1166, GB-1107, and Enzalutamide in Rat Plasma and Its Application to Pharmacokinetic Studies.Drug design, development and therapy · 2026
- Simultaneous determination of bosutinib and its metabolite in rat plasma by UPLC-MS/MS: Development, validation and application to pharmacokinetic and metabolic stability studies.Chemico-biological interactions · 2026
- Development and ICH M10-compliant validation of an LC-MS/MS method for quantification of atirmociclib in mouse plasma and its application to preclinical pharmacokinetic studies.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026