Clinical Chemistry

B-399 Lessons Learned: Reducing False-Positive Fentanyl Screens in a Pediatric Hospital Through Assay Selection

Xu Li, Luke Matthews, Bhavina Patel, Cristina Interiano, Van Leung-Pineda, Hiral Patel

Review

Key points

From the paper's conclusion
  • Recommended assay selection per colleagues, post–live monitoring, and diligent communication between the laboratory and pediatric providers are essential to ensure accurate fentanyl screening and appropriate clinical interpretation.

Abstract

Background Fentanyl use in pediatric patients has increased in recent years for both severe acute pain management and intensive care analgesia and sedation. As utilization grows, the need for a timely and accurate fentanyl detection has become increasingly important to our clinical providers. At our institution, we implemented Immunalysis fentanyl immunoassay successfully after validation. Unexpected positive fentanyl results were identified within the month in the patients without documented fentanyl administration or prior substance use history. In response, we promptly evaluated and transitioned to ARK Diagnostics Fentanyl II assay. This study summarizes the performance of both assays in a pediatric setting, with particular focus on false-positive rates based on LC-MS/MS confirmation and medical record review. Methods We conducted a retrospective observational study of all pediatric fentanyl immunoassay tests performed at our institution between 4/1/2025 and 1/21/2026. Two time periods were compared: (1) Immunalysis period (4/1/2025–10/22/2025), and (2) ARK period (10/23/2025–1/21/2026), following implementation of the replacement assay. Both assays were validated per the recommendation from the vendors. Meet requirements for accuracy, precision, reference intervals (cutoff), and patient comparison. For each period, fentanyl immunoassay results were extracted and correlated with electronic medical records for documented fentanyl administration and substance use history. False positives were defined as positive screening results with negative confirmatory testing and/or no documented fentanyl administration during the hospitalization or previous history of substance use. False-positive rates, overall positivities, and test utilization between assays were compared between assays. Selected suspected false-positive specimens from Immunalysis were stored and reanalyzed during ARK validation. Results During the Immualysis period, 3,007 specimens were tested, with an immunoassay positivity rate of 4.8%. Among positive screens, 25.5% (37/145) were false positives, yielding an adjusted positivity rate of 3.58%. In the ARKperiod, 1,391 specimens were tested, with a positivity rate of 3.52%. Test utilization was similar (15.4 vs 14.7 specimens/day Immualysis and ARK, respectively). 23 Immunalysis-positive specimens (13 suspected false positives and 10 expected positive per records) were verified with LC-MS/MS within 3 days. All 13 suspected false positives were negative by LC-MS/MS. Chart review showed that these 13 patients had been prescribed at least one of the following medicines: trazodone, haloperidol and risperidone. During ARK validation, 25 Immualysis –positive specimens (17 confirmed positives, 8 LC-MS/MS confirmed false positives) were re-tested. All 8 suspected false positives were negative on ARK. On repeat testing with Immunalysis after storage, 5 of these 8 were also negative, suggesting instability of the interfering substance(s) in urine. Conclusion In a pediatric hospital setting, the choice of fentanyl immunoassay had a substantial impact on the patient care because of the troubling false-positive rates. Immualysis assay showed significant cross reactivity with trazodone, haloperidol and risperidone. Replacement with ARK assay maintained similar overall positivity and utilization while reducing suspected false positives. Recommended assay selection per colleagues, post–live monitoring, and diligent communication between the laboratory and pediatric providers are essential to ensure accurate fentanyl screening and appropriate clinical interpretation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers