B-371 Streamlining Immunosuppressant Quantification: Evaluation of Automated Magnetic Bead–Based Solid-Phase Extraction (mSPE) for Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)

Yi Xiao, Le Meng, Q I an Cheng, Edward Leung

Key points

From the paper's conclusion
  • Preliminary evaluation for everolimus and cyclosporine A showed mean biases of 6.2% and 6.4%, respectively, with additional patient comparison studies ongoing.

Abstract

Background Reliable sample preparation is the cornerstone of robust clinical mass spectrometry. An optimal method must effectively remove biological matrix components and potential interferences while minimizing hands-on time and operational complexity. While protein precipitation is widely adopted for its simplicity, it may fail to effectively remove complex matrix components such as lipids, leading to significant matrix effects (i.e., ion suppression or enhancement) that may contribute to increased instrument maintenance and downtime. Conversely, conventional cartridge- or plate-based Solid Phase Extraction (SPE) offers superior cleanup but relies on uniform flow-through sorbent beds, which may be prone to clogging and increase mechanical complexity. Magnetic bead–based SPE (mSPE) overcomes these limitations by providing selective cleanup via suspension-based magnetic handling, enabling high-throughput, "walk-away" automation, with the potential to enable shorter analytical methods and improved turnaround time. We describe the evaluation of the Xpeedy® mSPE technology (Cura Diagnostics) for Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS) quantification of tacrolimus, sirolimus, everolimus, and cyclosporine A. Methods Whole blood specimens (15 µL) were lysed and mixed with internal standards in a 96-well format. Extraction was performed using the KingFisher Flex system (ThermoFisher) with a 20-minute automated mSPE protocol involving sequential wash and elution steps. Following centrifugation, 5 µL of eluate was analyzed on a ACQUITY UPLC I-Class PLUS (Waters) coupled to a TQ Absolute mass spectrometer (Waters) using a 2.0-minute gradient with an ACQUITY HSS C18 SB column (1.8 µm, 2.1 × 30 mm, Waters). Repeatability, reproducibility, and carryover were evaluated using Quality Control (QC) materials. Analytical Measuring Range (AMR) was established. Matrix effect and recovery were quantified via the Matuszewski method. Matrix effect was further evaluated using a matrix dilution study. Accuracy was evaluated against a 5-minute validated in-house LC-MS/MS method. Results Within-run and between-run imprecision were < 9% CV for all four immunosuppressants, with AMRs appropriately spanning therapeutic ranges. The mSPE technology yielded high extraction recoveries (76 – 92%) and negligible matrix effects (within ± 10% relative to neat solution response). This enhanced cleanup supported a rapid 2-minute LC-MS/MS method. Method comparison with the in-house LC-MS/MS assay demonstrated strong agreement for tacrolimus and sirolimus. Passing-Bablok regression yielded a slope of 1.06 and an intercept of 0.16 ng/mL for tacrolimus, and a slope of 1.15 and an intercept of –0.31 ng/mL for sirolimus. Mean biases were 7.8% and 12.4%, respectively. Preliminary evaluation for everolimus and cyclosporine A showed mean biases of 6.2% and 6.4%, respectively, with additional patient comparison studies ongoing. Conclusion The mSPE technology provides a robust and scalable alternative to conventional sample preparation methods. By combining the selectivity of SPE with suspension-based magnetic handling, the assay minimizes matrix effects, improves workflow efficiency, and supports high-throughput clinical therapeutic drug monitoring of immunosuppressants.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers