B-313 Validation of the MGISP-100 Automation Platform for Pre-Capture NGS Library Preparation in Clinical Workflows
Gustavo Barra, Pedro Mesquita, Milena Torres, Anderson Santos, Nara Pessoa, Ilária Sgardioli, Amanda Lamas, Lorena Gama,
Key points
From the paper's conclusion- Overall, MGISP-100 supports standardized, scalable pre-capture library preparation suitable for routine clinical NGS implementation without compromising key QC or sample-integrity indicators.
Abstract
Background Automation of next-generation sequencing (NGS) library preparation can improve efficiency, reproducibility, and contamination control by standardizing pipetting-intensive steps and reducing operator-dependent variability. In clinical laboratories, these gains may translate into more predictable turnaround times, reduced hands-on labor, and tighter control of process critical quality attributes (CQAs) that impact downstream sequencing performance. Here, we evaluated the MGISP-100 Automated Sample Preparation System (MGI Tech) for pre-capture library preparation in a routine-like clinical setting, using CQAs and sample-integrity indicators relevant to day-to-day laboratory release criteria. Methods Fifty-three routine-like specimens (29 whole-blood EDTA and 24 oral swabs; 24 males and 29 females) were processed. Genomic DNA was extracted using Maxwell (Promega). Pre-capture library preparation (DNA fragmentation, end-repair/A-tailing, adapter ligation, SPRI-based cleanups, and library PCR amplification) was performed either by an established manual workflow or on the MGISP-100 Automated Sample Preparation System using the xGen DNA Library Prep kit and xGen UDI Primer Plate 1 (Integrated DNA Technologies). Libraries were quantified by Qubit (Thermo Fisher Scientific) and sized by TapeStation 4200 (Agilent). Capture/enrichment was held constant (manual, identical conditions in both arms) and was not evaluated as an outcome; analyses focused on pre-capture library preparation. Sequencing was performed on DNBSEQ-G400 (paired-end 300-cycle). A reference control sample was prepared under both conditions for head-to-head variant calling comparison using DRAGEN v4.3.17 (Illumina), summarized by F-measure across SNVs, INDELs, and CNVs. To assess sample integrity and potential mixture/cross-contamination, IDR_BAF (inter-decile range of B-allele frequency across variant sites) was extracted from validation-run samples and compared between workflows. Comparisons were reported overall and stratified by specimen type (whole blood EDTA vs swab). Results No MGISP-100–specific failures occurred, and QC pass rates were consistent across specimen types. Libraries met sequencing QC targets with appropriate fragmentation and yield (mean fragment size: MGISP-100 342.76 ± 13.19 bp vs manual 334.20 ± 20.80 bp; mean concentration: MGISP-100 70.97 ± 12.81 ng/µL vs manual 64.33 ± 13.97 ng/µL). In the control sample, variant calling performance was comparable between workflows (F-measure: SNVs > 0.97, INDELs > 0.77, CNVs > 0.50). IDR_BAF values remained below the laboratory alert threshold (<0.20) in both groups, supporting preserved sample integrity. Overall mean IDR_BAF was higher with MGISP-100 (0.12328 ± 0.01096; n=25) than manual preparation (0.11464 ± 0.00776; n=28) (p=0.00210; 95% CI 0.00332–0.01396), while remaining below the threshold. Conclusion In both whole-blood EDTA and swab specimens, the MGISP-100 Automated Sample Preparation System produced pre-capture libraries with QC metrics and downstream variant calling performance comparable to the manual workflow. The small shift in IDR_BAF observed with MGISP-100—although statistically significant—remained below the established alert threshold and is consistent with minor process effects (e.g., subtle differences in pipetting dynamics, bead mixing, or cleanup stringency) rather than actionable cross-contamination. Overall, MGISP-100 supports standardized, scalable pre-capture library preparation suitable for routine clinical NGS implementation without compromising key QC or sample-integrity indicators.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An Automated Microfluidic NGS System for Sensitive and Standardized Detection: A Comprehensive Analytical and Clinical Validation.The Journal of molecular diagnostics : JMD · 2026
- Microfluidic approaches to next- generation sequencing library preparation: innovations, clinical integration and point-of-care settings.Expert review of molecular diagnostics · 2026
- Robotic Centrifugal Microfluidic Automation Enables Library Preparation of Nine Samples with a Tenfold Reduction in Reagent Consumption for Decentralized Sequencing.Analytical chemistry · 2026
- A Framework for Analytical Validation of Genomic DNA-Based Shotgun Metagenomic Next-Generation Sequencing WorkflowsmedRxiv · 2026
- Automated all-in-one device for handling Magnetic Bead-Based Samples in immunoassays.Device · 2026
- Comparative performance of portable DNA extraction protocols and bioinformatics workflows for rapid detection of gram-negative bacteria and antimicrobial resistance using Oxford Nanopore sequencing.Microbiology spectrum · 2026