B-284 Validation of Cystatin C on Abbott Alinity C

Ka Hang Brian Lam, Abelyn Villamor, Semee Park, Dominique Ramos, Gene Usher, Vincent Dizon, Darryl Palmer-Toy

Key points

From the paper's conclusion
  • Abbott has recently released its FDA-approved cystatin C reagent in the United States, and here we present its validation on the Alinity C.

Abstract

Background For decades, creatinine and the associated estimated glomerular filtration rate (eGFR) calculation have been the standard for the assessment of kidney function. However, there are a number of limitations, including the association with muscle mass, race, and staging of chronic kidney disease. An alternative to creatinine is cystatin C, which is a protein produced at a constant rate by nucleated cells. As such, it does not suffer from many of the limitations of creatinine, and many studies show improved risk stratification for death, cardiovascular disease, and end-stage renal disease. Abbott has recently released its FDA-approved cystatin C reagent in the United States, and here we present its validation on the Alinity C. Methods Serum samples with creatinine measurements were used to obtain cystatin C measurements across the analytical measurement range. The Beckman gentian linearity reagent was purchased to perform linearity and accuracy studies. Correlation studies were performed against an outside reference laboratory. Results Forty serum samples were run using the Abbott Alinity C. Precision studies were acceptable against a total allowable error of 10%. Correlation studies against an outside reference laboratory showed a negative bias. Conclusion The Abbott assay can be used in clinical laboratories to measure cystatin C. As with all assays, there are biases, and laboratory personnel should be well aware of these before adopting a new assay.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers