Clinical Chemistry

B-207 Increasing Diagnostic Specificity Without Sacrificing Sensitivity in IgE testing

Julia Sinner, Peter Forstenlechner, Christian Lupinek, Philipp Matern

Case report

Key points

From the paper's conclusion
  • These findings were consistent with the clinical symptoms and the SPT results, effectively excluding clinically relevant pollen sensitizations.

Abstract

Background Cross-reactive Carbohydrate Determinants (CCDs) are carbohydrate epitopes found in plants, insect venoms, and mites. Approximately 20% of allergic patients generate anti-CCD IgE antibodies which are clinically irrelevant and can cause "false-positive" results in sIgE-testing. A 13-year-old male patient presented with recurrent hives of unclear origin and no respiratory symptoms. Initial diagnostic workup revealed a significant discrepancy: while extract-based sIgE testing suggested a broad sensitization to various respiratory allergens, Skin Prick Test (SPT) results were negative. This marked discordance between the patient’s clinical history, SPT results, and sIgE data necessitated advanced molecular allergen testing and CCD inhibition to resolve the diagnostic gap. Methods To investigate potential CCD interference, the serum was referred for secondary analysis. Testing included:ImmunoCAP: following pre-incubation with a CCD inhibitor, sera were tested for sIgE to the same extracts used for initial diagnostic workup (Birch, Timothy Grass, Ash, Ribwort, Mite mix) and to the CCD marker MUXF3.Allergy Xplorer 3 (ALEX³): Analysis via the ALEX³ multiplex test, which measures sIgE against 300 extracts and molecular allergens, utilizing its integrated CCD inhibition protocol. Results Initial unblocked ImmunoCAP testing demonstrated broad extract reactivity: Timothy grass (17.2 kUA/L), mite mix (1.51 kUA/L), birch (1.78 kUA/L), ash (1.98 kUA/L), ribwort (1.94 kUA/L), and the CCD marker MUXF3 (4.44 kUA/L). Following pre-incubation with the CCD inhibitor, values markedly decreased. Birch, ash, ribwort, and MUXF3 converted to negative (<0.35 kUA/L). Timothy grass sIgE dropped from 17.2 kUA/L to 0.92 kUA/L (>94% reduction), indicating that the initial value was driven primarily by CCD interference rather than true protein sensitization. In parallel, ALEX³ with its by default integrated CCD inhibitor showed negative results for the respective allergens used as markers for specific sensitisation to timothy grass, birch, ribwort and mites (<0.10 kUA/L) and the CCD marker Hom s LF remained below the threshold of 0.3 kUA/L (0.19 kUA/L). These findings were consistent with the clinical symptoms and the SPT results, effectively excluding clinically relevant pollen sensitizations. Conclusion The presence of CCDs remains a significant hurdle in allergy diagnostics, frequently triggering misleading, non-specific sIgE results. As demonstrated in this case, CCD inhibition is an essential requirement for the accurate interpretation of sIgE profiles. By resolving these diagnostic discrepancies, ALEX³ helps to prevent the risk of misdiagnosis and the subsequent prescription of unnecessary allergen avoidance or inappropriate allergen immunotherapy.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers