A-403 Verification of the Low-Range Activated Clotting Time Test on the GEM Hemochron 100

Antonio Trammel

Key points

From the paper's conclusion
  • The objective of this study was to verify the performance specifications of the low range activated clotting time (ACT-LR) test, used to monitor low to moderate UFH doses, on the GEM100 as compared to the Signature Elite.

Abstract

Background The activated clotting time (ACT) test is commonly used to monitor unfractionated heparin (UFH) anticoagulation therapy during cardiac procedures. The GEM Hemochron 100 (GEM100) and its predicate device, the Hemochron Signature Elite, are point of care (POC) analyzers which utilize identical test cartridges and controls to perform ACT testing. Updates with the newer GEM100 device include wireless, bidirectional connectivity to streamline reporting. The objective of this study was to verify the performance specifications of the low range activated clotting time (ACT-LR) test, used to monitor low to moderate UFH doses, on the GEM100 as compared to the Signature Elite. Methods The ACT-LR test was evaluated on three new GEM100 devices. Intra-assay precision was assessed by testing 2 levels of quality control (QC) 10 consecutive times within the same day, while inter-assay precision was evaluated by testing 2 levels of QC twice a day for 5 days. Method comparison was performed using whole blood samples from 45 patients to evaluate analytical agreement between the Signature Elite and the GEM100. Accuracy was further assessed by comparing inter-assay QC means with manufacturer-assigned target means. As manufacturer’s reference intervals are published for both normal donor and patient populations, samples from both groups were evaluated for verification studies. Analysis was performed by EP Evaluator. Acceptance criteria are detailed in Table 1. Results Intra- and inter-assay precision were within acceptable limits with a highest coefficient of variation (CV) for level 1 at 8.7% and 7.5%, respectively, and for level 2 at 7.7% and 8.5%, respectively. Compared to the manufacturer’s target means, % bias was at most 7.7% for level 1 and 3.2% for level 2. All GEM100 devices demonstrated acceptable analytical agreement with the Signature Elite, with slopes ranging from 0.97-1.01, correlation coefficients (R) all >0.97, and % mean bias at most 6.7%. The manufacturer’s normal healthy donor reference interval (116-155 Celite equivalent seconds) could not be verified, with 15.4% of values falling outside of this range. The manufacturer’s reference range for patients not on heparin therapy (83.2-199.5 Celite equivalent seconds), however, was verified by testing the intended patient population. Conclusion Testing of the ACT-LR cartridge on the GEM100 demonstrated acceptable performance and analytical agreement with the Signature Elite. While two sets of reference ranges are provided by the manufacturer, the healthy donor range proved to be too narrow for our population and only the non-UFH patient range could be verified. Given the clinical use of the ACT test to monitor patients undergoing procedures requiring UFH treatment, the manufacturer’s range for this population may be more clinically meaningful and was adopted at our institution.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers