A-364 External Quality Assessment and Proficiency Testing Performance of a Prototype Multiplexed Sample-To-Answer Diagnostic System for Identification of Vaginal Infections from Vaginal Swabs

Monica Cronin, Crystal Smith, Ashley Hillman, Jessica Papenfuss, Brad Graham, Margarita Rogatcheva

Key points

From the paper's conclusion
  • One missed detection of C. albicans (QCMD CAND21S-03) may be attributed to organism concentration below the assay’s limit of detection (LoD), consistent with <80% detection of this organism across other diagnostic platforms.

Abstract

Background bioMérieux is developing the BIOFIRE® SPOTFIRE® Vaginitis (VG) Panel, a syndromic, molecular test designed for the point-of-care setting. The Panel prototypes test vaginal swab specimens collected in eNAT® media and provides rapid results (approximately in 20 minutes) for Bacterial Vaginosis (BV), Vulvovaginal Candidiasis (VVC) (with specification of six Candida species), and trichomoniasis. Proficiency Test (PT) and External Quality Assessment (EQA) programs provide important quality management tools for the clinical laboratory, supporting the accuracy and reliability of patient testing. Single and multi-pathogen EQA and PT programs were evaluated using Research-Use-Only (RUO) and Investigational-Use-Only (IUO) prototypes of the BIOFIRE® SPOTFIRE® VG Panel to assess performance across PT and EQA challenges from 2018-2025. Methods PT and EQA programs intended for molecular detection of BV, Candida species and Trichomonas vaginalis were tested using RUO and IUO BIOFIRE® SPOTFIRE® VG Panels on the BIOFIRE® SPOTFIRE® System. PT programs from the American Proficiency Institute (API), the College of American Pathology (CAP), Wisconsin State Laboratory of Hygiene (WSLH), and EQA programs from Quality Control for Molecular Diagnostics (QCMD) were tested as directed by the provider and results were compared to published reports. Results BIOFIRE® SPOTFIRE® VG RUO and IUO Panel test results for 205 PT/EQA samples from 18 different surveys demonstrated 14/14 (100%) expected detections of BV, 64/64 (100%) detections for T. vaginalis, and 93/94 (98.9%) detections of Candida organisms associated with VVC: C. albicans: 46/47 (97.9%), C. dubliniensis: 1/1 (100%), C. glabrata: 22/22 (100%), C. krusei: 16/16 (100%), C. parapsilosis: 7/7 (100%), and C. tropicalis: 1/1 (100%). PT/EQA challenges intended to be BV negative were correctly reported in 52/52 samples (100%) and challenges intended to be negative for VVC and T. vaginalis were concordant in 39/39 samples (100%). A single unexpected detection of C. parapsilosis occurred in WSLH 2025 VPM-2; repeat testing could not be performed due to loss of sample viability. One missed detection of C. albicans (QCMD CAND21S-03) may be attributed to organism concentration below the assay’s limit of detection (LoD), consistent with <80% detection of this organism across other diagnostic platforms. Conclusion These results demonstrate that the BIOFIRE® SPOTFIRE® VG Panel can accurately identify BV and vaginal pathogens across multiple PT/EQA panels while providing a rapid sample to answer solution suitable for point-of-care molecular diagnostics. This abstract contains data regarding a device that has not been reviewed or approved by regulatory agencies for in vitro diagnostic use.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers