A-354 Evaluating the Analytical Accuracy of NBCL Intraoperative PTH POCT Versus the Abbott Alinity Chemiluminescent Immunoassay

Waleed Tamimi, Abdulrahman Alanazi, Rehaf Alabdaly, Najla Alhussain, Alanoud Alharbi, Alanood Alhussainan, Bashayer Alenazi, Nawaf Almutairi,

Key points

From the paper's conclusion
  • Therefore, intraoperative monitoring of intact PTH (iPTH) levels plays a critical role in confirming the complete removal of hypersecreting tissue and minimizing the risk of persistent or recurrent disease.[2] The use of Point-of-care testing (POCT) devices for intraoperative ioPTH measurement can offer rapid turnaround time by producing results within minutes and facilitating real-time surgical d

Abstract

Background Primary hyperparathyroidism is a common endocrine disorder characterized by the excessive secretion of parathyroid hormone (PTH), often due to parathyroid adenomas.[1] Surgical excision of the hyperfunctioning gland remains the definitive treatment. Therefore, intraoperative monitoring of intact PTH (iPTH) levels plays a critical role in confirming the complete removal of hypersecreting tissue and minimizing the risk of persistent or recurrent disease.[2] The use of Point-of-care testing (POCT) devices for intraoperative ioPTH measurement can offer rapid turnaround time by producing results within minutes and facilitating real-time surgical decision-making.3 These devices aim to provide surgeons with immediate feedback on the success of the procedure, reducing the need for extended operative time and potential re-exploration.[2] The accuracy of POCT devices is often questioned compared to gold-standard central laboratory methods like the Abbott Alinity i, as result differences can impact surgical decisions and patient outcomes. This study aims to compare the analytical accuracy of the Novel Biomarkers Catalyst Lab (NBCL), Netherlands ioPTH POCT device against the Abbott Alinity, USA, method, assessing their agreement and potential interchangeability in the surgical management of hyperparathyroidism. The aim of this study was to compare and evaluate the ioPTH level results accuracy of NBCL Point of care ioPTH with that obtained from Abbott Alinity i immunoassay in the Central Laboratory. Methods This study enrolled 9 patients scheduled for parathyroidectomy surgery and obtained 30 venous samples in SST tube and 30 whole blood samples in EDTA tube. Whole Blood venous samples were collected simultaneously from each participant for measuring of ioPTH using point-of-care devices (NBCL) and sent to the central laboratory instrument (Abbott Alinity i). Samples were run intraoperative as follows: Pre-operative, Pre- excision, post-excision 5 minutes, post-excision 10 minutes, post-excision 20 minutes, post-excision 30 minutes. The venous samples were collected by anesthesiologist during the operation and sent to the laboratory immediately after sample collection. Results For the method comparison, analysis was based on the CLSI EP09-A2 protocol Deming and regular regression analysis was performed to compare the NBCL Intraoperative Intact PTH (ioPTH POCT Device) against the laboratory method Abbott Alinity i.[3] The Evaluation Criteria included the Allowable Total Error: TEa ±30% (CLIA, WSLH, AAB, API) CLIA: CLIA – Federal Register, WSLH: Wisconsin State Laboratory of Hygiene; AAB: American Association of Bioanalysts; API: American Proficiency Institute. The Correlation Coeff (r)= 0.9866 with slope=0.9385 and intercept=1.4169 at TAE of 30%. The total agreement between the two analyzers was found to be 80%. This agreement was reduced to 66.7% at lower levels below 11 pg/ml, however, it was improved to 89% above this cutoff. The majority of disagreement values were observed in the post-excision time (75%). Conclusion The PTH results appeared to be influenced by surgical manipulation reaching a peak then a steady decline post extraction. As a result of the study, iPTH had a good comparison agreement and pass Accuracy-Correlation with reference laboratory and clinical acceptance from the surgeons that the results have met their expectations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers