A-333 Analytical performance evaluation of VITROS™ Immunodiagnostic Products Progesterone 2 Assay

Qian Ding, Rea Castro, Jody Parsells, Jeff Houtz, Jodiann Atienza, Scott Hewitson

Key points

From the paper's conclusion
  • Conclusion The VITROS Progesterone 2 Assay demonstrated acceptable analytical performance across evaluated parameters including precision and limit of quantitation, with very high correlation and close agreement to the Roche Elecsys Progesterone III assay.

Abstract

Background Progesterone can be used as an aid in the diagnosis and treatment of disorders of the ovaries or placenta. Designed with a sheep monoclonal antibody rather than a polyclonal rabbit antibody, the VITROS Immunodiagnostic Products (VITROS Progesterone 2) Assay is a next-generation immunoassay developed to improve analytical specificity through key reagent redesign to replace the VITROS Immunodiagnostic Products Progesterone Assay. The assay employs a sheep monoclonal antibody, biotinylated and overcoated onto streptavidin-coated wells to mitigate susceptibility to biotin interference. Studies were conducted to evaluate the analytical performance of the VITROS Progesterone 2 Assay by determining limit of quantitation (LoQ) and precision. A method comparison study of the VITROS Progesterone 2 Assay with the Roche Elecsys Progesterone III assay was also conducted. Methods The LoQ was determined consistent with CLSI document EP17. Precision was determined consistent with CLSI EP05. Repeatability (within-run) percent coefficient of variation (%CV) was determined using two replicates per run. Reproducibility (within-lab) %CV was determined using a single reagent lot and a single calibration. A quantitative comparison between the VITROS Progesterone 2 Assay and the Roche Elecsys Progesterone III assay was performed in accordance with CLSI guideline EP09 using 125 serum samples. The VITROS Progesterone 2 Assay was screened for interfering substances at progesterone concentrations of approximately 3.00 ng/mL, and 12.0 ng/mL following CLSI EP07 12 and EP37. Results The claimed LoQ was determined using a precision profile approach with a %CV limit of 20% and set at 0.335 ng/mL. Precision was observed at low, intermediate and high regions of the measuring interval. At a progesterone concentration of approximately 0.9 ng/mL, repeatability was 4.5%CV and reproducibility was 6.4%CV. At a progesterone concentration of approximately 19.5 ng/mL, repeatability was 1.7%CV and reproducibility was 2.5%CV. At a progesterone concentration of approximately 35 ng/mL, repeatability was 1.8%CV and reproducibility was 2.7%CV. The method comparison between the VITROS Progesterone 2 Assay and the Roche Elecsys Progesterone III assay yielded a correlation coefficient of 0.995 and a Passing-Bablok slope of 0.99 and intercept of -0.47. Interference testing demonstrated no biotin interference effect on assay performance at concentrations up to 3,510 pg/mL. Conclusion The VITROS Progesterone 2 Assay demonstrated acceptable analytical performance across evaluated parameters including precision and limit of quantitation, with very high correlation and close agreement to the Roche Elecsys Progesterone III assay.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers