A-330 Analytical performance evaluation of the Diazo Total Bilirubin assay on the Atellica CH Analyzer for use in neonates
Mark Borgman, Dejan Blagovcanin, Zachary Eischeid, Anne Sinopoli, Jennifer Cheek, Nazira Ozgen, Jennifer Snyder
Key points
From the paper's conclusion- Results The Atellica CH D_TBil assay demonstrated agreement with the Roche cobas TBIL assay across the neonatal concentration range of 0.10–21.81 mg/dL, with a regression of y = 1.06x - 0.02 mg/dL and r = 0.997, indicating strong analytical concordance within clinically relevant bilirubin levels.
Abstract
Background Total bilirubin (TBil), a byproduct of hemoglobin catabolism, is a critical biomarker in the evaluation of neonatal hyperbilirubinemia. In newborns, even modest elevations in bilirubin levels are closely monitored due to the risk of kernicterus, a serious and potentially irreversible neurological complication associated with excessive bilirubin exposure. Accurate measurement of TBil is therefore essential for early identification, risk stratification, and clinical management of hyperbilirubinemia in neonates. This study evaluated the reference range and method comparison of the Diazo TBil (D_TBil) assay on the Atellica CH analyzer using prospectively collected neonatal serum samples. Methods Reference Interval: Healthy neonatal serum samples with no history of hyperbilirubinemia were prospectively collected (23 samples: 0–1 day; 27 samples: 1–2 days; 26 samples: 3–5 days). All samples were tested using the Atellica CH D_TBil assay. Results were compared with published reference ranges, and ranges were considered verified if =10% of the samples fell outside of the established limits. Method Comparison: A total of 113 prospectively collected serum samples from healthy neonates and those diagnosed with hyperbilirubinemia, along with 11 contrived samples prepared by spiking neonatal serum with unconjugated bilirubin, were tested using the Atellica CH 930 D_TBil and Roche cobas 8000 TBIL (BILT3) assays. Weighted Deming regression was used to assess slope, y-intercept, and correlation between the two methods. Results The Atellica CH D_TBil assay demonstrated agreement with the Roche cobas TBIL assay across the neonatal concentration range of 0.10–21.81 mg/dL, with a regression of y = 1.06x - 0.02 mg/dL and r = 0.997, indicating strong analytical concordance within clinically relevant bilirubin levels. The Atellica CH D_TBil assay demonstrated complete concordance with physician-assigned classifications of healthy versus hyperbilirubinemic neonates when evaluated against the established clinical nomogram (Kemper et al., Pediatrics. 2022 Sep 1;150(3)). Age specific expected TBil values were verified for healthy neonates aged 0-1 day (<8.00 mg/dL), 1-2 days (<12.00 mg/dL), and 3-5 days (<16.00 mg/dL), consistent with clinical expectation. Conclusion Using neonatal serum samples, the Atellica CH D_TBil assay demonstrated clinically acceptable agreement with the Roche cobas TBIL assay and verified age-appropriate neonatal reference values, supporting its use for accurate assessment and management of neonatal hyperbilirubinemia.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detection of individual false-low total bilirubin results in neonates and young infants using the bilirubin-to-icteric index ratio.Clinical biochemistry · 2026
- Comparability of Total Bilirubin methods for neonatal samples across 6 NHS trusts within North West England.Annals of clinical biochemistry · 2026
- Reference intervals for plasma total bilirubin in children.Scandinavian journal of clinical and laboratory investigation · 2026
- Commutability assessment of candidate reference materials for total bilirubin assays in neonates.Clinical chemistry and laboratory medicine · 2026
- Measuring total bilirubin for clinical decision making: are we treating at the right number?Seminars in perinatology · 2026
- Case Report: Usefulness of transcutaneous bilirubin measurement for the early detection of severe hyperbilirubinemia due to late-onset hemolysis in an extremely low birth weight infant.Frontiers in pediatrics · 2026