A-184 Cystic Fibrosis Screening: Modernizing Sweat Collection and Chloridometry

Olumuyiwa Adesoye, Makenzie Breiter, Melinda Rankin, Joanna Bojarski, Lisa Johnson

Key points

From the paper's conclusion
  • The paired sample T-test suggested there was no significant difference between the volume of sweat collected between the 3700 and the Advanced sweat collectors.

Abstract

Background Cystic Fibrosis (CF) is caused by mutations in the cystic fibrosis transmembrane conductance regulator (CFTR) gene, resulting in defective, insufficient, or absent CFTR proteins. As these proteins function as critical chloride ion channels, their dysfunction leads to the sequestration of chloride ions within epithelial cells and a subsequent reduction in extracellular fluid. This manifests clinically as dehydrated, viscous mucus accumulation on epithelial surfaces and impaired mucociliary clearance. The quantitative measurement of sweat chloride is a diagnostic test for CF. Historically, many laboratories have utilized general-purpose automated chloridometers, such as the Nelson-Jameson Chloride Analyzer, which were originally designed for food-grade salinity analysis. Our laboratory sought to modernize this workflow by transitioning to a platform specifically engineered for the clinical environment: the ChloroChek® Chloridometer and Macroduct® Advanced system. Methods We conducted a technical assessment of the Macroduct® Advanced system (Model 3710) in comparison to the legacy Macroduct® 3700 and Nelson-Jameson analyzer. The validation protocol included a within-run precision study (10 replicates of three control levels, L1-L3) and a between-run study (20 runs over 10 days). Accuracy was evaluated via method comparison using 25 sweat specimens spanning the reportable range with an acceptability criteria of R^2 = 0.95 and slope 0.90-1.10. Analytical Measuring Interval (AMI) validation was performed using five linearity samples run in triplicate, targeting an R^2 =0.99 and recovery within ±5% or 2 mmol/L. Our evaluation compared the performance of the 3700 and Advanced sweat collectors by performing single-arm collections with each device on the same individual; the paired sample T-test was used to assess differences in volume of collected sweat (p=0.05). Additionally, we conducted studies to assess sample stability, carryover, and reference ranges. Results Method comparison against the Nelson Jameson legacy system demonstrated strong correlation with a Deming regression of y = 0.91x + 2.2 and R^2 = 0.99. Within-run precision for two ChloroChek® Chloridometers yielded mean values of 22.1–22.9 mmol/L (L1), 46.2–46.8 mmol/L (L2), and 96.1–97.7 mmol/L (L3). All replicates demonstrated a coefficient of variation (CV) < 5% (1.25%–4.34%). AMI validation (10–160 mmol/L) showed excellent linearity (R^2 = 0.99) and recovery within ±5% of target values. The paired sample T-test suggested there was no significant difference between the volume of sweat collected between the 3700 and the Advanced sweat collectors. Conclusion Our data indicate acceptable analytical performance. Transitioning to an FDA-cleared, purpose-built clinical system represents a significant advancement in CF diagnostics. Our laboratory will utilize a phased soft launch approach, initiating testing with the adult population before expanding to pediatric and neonates.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers