A-084 Performance of the BD Enteric Bacterial Panel plus for BD COR™ System
Ray McMillian, Stephany Ma, William Harty, Sonia Paradis
Key points
From the paper's conclusion- Conclusion The BD EBP plus assay for BD COR System provides rapid, accurate detection of a broad range of enteric bacterial pathogens, demonstrating strong analytical and clinical performance.
Abstract
Background In 2025, Becton Dickinson received FDA 510(k) clearance (K250358) for the BD Enteric Bacterial Panel (BD EBP) and BD EBP plus assays on the BD COR™ System, automated qualitative in vitro diagnostic tests for enteric bacterial pathogens. BD EBP detects Salmonella spp., Campylobacter spp., Shigella spp./enteroinvasive Escherichia coli (EIEC), and Shiga toxin (stx1/stx2) genes, while BD EBP plus additionally detects Plesiomonas shigelloides, Vibrio spp., enterotoxigenic E. coli (ETEC) LT/ST genes, and Yersinia enterocolitica. Analytical and clinical performance of BD EBP plus are described. Methods Analytical sensitivity (LoD) was established using two quantified representative strains per target at multiple concentrations in pooled stool matrix preserved in BD Fecal Collection and Transport tubes, tested across three BD COR Systems and three reagent lots,. Site to site (STS) reproducibility was evaluated at three clinical sites using one reagent lot, and lot to lot (LTL) reproducibility at one internal site using three reagent lots; each study included 60 runs. Cross contamination was assessed using high positive ETEC samples (=1 × 106 CFU/mL) alternated with negative samples across 18 runs on three BD COR systems. Potential interference was evaluated using 26 biological and chemical substances and 14 endogenous microorganisms. The clinical performance of BD EBP plus on the BD COR System was evaluated in a multi-site study across six geographically diverse clinical centers using stool specimens collected during routine care. A total of 1,493 prospective and 200 retrospective samples were evaluable. Positive and negative percent agreements (PPA/NPA) were determined using a single FDA-cleared comparator assay for Salmonella spp., Shigella spp./EIEC, and P. shigelloides, or a two-of-three comparator method for all other targets. Contrived specimens supplemented low-prevalence targets. Results LoDs ranged from 116 (C. coli) to 10,705 units/mL (V. vulnificus) across targets. STS and LTL reproducibility demonstrated 100% agreement for true negatives, 98.9%–100% agreement for low positive, and 97.8%–100% for moderate-positive samples. The Ct coefficients of variation ranged from 1.0% to 3.4% for STS. No false positives were observed in cross- contamination testing. Three substances displayed interference: Nystatin at concentrations above 25%; hydrocortisone, above 15%; and Imodium, above 2.5 mg/mL. Only E. coli (stx1/stx2) and Bacteroides vulgatus displayed interference at very high concentrations (>2.0 × 106 CFU/mL). PPA and NPA for prospective and retrospective specimens combined were calculated for each target in the BD EBP plus assay. PPA was >90% for all targets except P. shigelloides and ETEC. PPA for P. shigelloides and ETEC was 80.0% (4/5) and 83.3% (25/30). NPA was >99% for all targets. Contrived specimens were used to supplement low-prevalence targets. PPA for contrived testing with the BD EBP plus assay was 100% across all target concentrations for all low prevalence organisms, except Vibrio spp., which obtained 96.0%, due to two false negative results at approximately 1.99× the LoD. Conclusion The BD EBP plus assay for BD COR System provides rapid, accurate detection of a broad range of enteric bacterial pathogens, demonstrating strong analytical and clinical performance. When used in appropriate clinical contexts, it supports timely, evidence-based management management of enteric infections.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic accuracy of a multiplex nucleic acid amplification test for detecting gastrointestinal bacterial pathogens: A multicenter clinical study.Diagnostic microbiology and infectious disease · 2026
- Multicentre Evaluation of Cepheid Xpert® Research-Use-Only Panel for Molecular Rapid Detection of Gastrointestinal Pathogens.Microorganisms · 2026
- Assessment of BioFire Gastrointestinal Panel unconfirmed positive results for enteropathogenic bacteria and Norovirus and accuracy of a post-analytical algorithm to mitigate unconfirmed results.Journal of clinical microbiology · 2026
- Phylogenetic Analysis of cpn60 UT region reveal oligonucleotide regions for the detection of prominent Enterobacteriaceae pathogensbioRxiv · 2026
- Phylogenetic analysis of the cpn60 UT region identifies conserved oligonucleotide regions for Enterobacterales detection.Brazilian journal of microbiology : [publication of the Brazilian Society for Microbiology] · 2026
- A novel cost-effective methodology for Campylobacter isolation from human stool and chicken sample matrix containing extended-spectrum β-lactamase-producing Escherichia coli.Journal of microbiological methods · 2026