A-069 Sensitivity of Antigen Testing for Histoplasmosis and Blastomycosis in Bronchoalveolar Lavage (BAL) and Cerebrospinal Fluid (CSF) Versus Urine and Serum
Dane Granger
Key points
From the paper's conclusion- Minimal increases in sensitivity were observed in BM/HP BAL versus U/S.
Abstract
Background Detecting Histoplasma (HP) and Blastomyces (BM) antigen (Ag) in urine (U) and serum (S) is an essential diagnostic tool for these pathogens. Labs have added bronchoalveolar lavage (BAL) and cerebrospinal fluid (CSF) as additional HP and BM Ag testing sources. We evaluated the sensitivity of Ag detection in U and S in patients with detected HP or BM Ag in BAL or CSF by enzyme immunoassay (EIA). Methods 14 patients with HP or BM Ag-positive results made up the CSF cohort. The Ag concentrations were below the limit of quantification (< LoQ, N=3), above the limit of quantification (> LoQ, N=1), and 0.3 to 9.5 ng/mL (mean = 2.3 ng/mL, N=10). The BM Ag-positive in BAL group included 24 patients, with concentrations of LoQ (N=5), and 0.3 to 11.3 ng/mL (mean = 3.5 ng/mL, N=15). HP Ag-positive in BAL comprised 36 patients, with 3 LoQ, and 23 ranging from 0.8 to 16.9 ng/mL (mean = 4.7 ng/mL). All patients were checked for routine clinical HP and BM EIA Ag test results on U and S. HP BAL patients were also investigated for cross-reacting galactomannan (GM; Aspergillus EIA in S or BAL) and (1?3)-ß-d-glucan (BDG; Fungitell EIA in S) antigen test orders. Test results greater than 14 days before or after CSF or BAL Ag testing were excluded. Results 100% (14/14), 95.7% (22/23), and 83.3% (30/36) of HP/BM CSF, BM BAL, and HP BAL Ag-positive patients were also positive in U and/or S, when both additional sources were ordered, respectively. Of the 6 HP BAL patients without positive U or S results, 3 had GM and BDG testing performed. All 3 had high signal GM and/or BDG results in S and/or BAL. In conjunction with low HP Ag levels in BAL (1.0 to 1.53), these findings suggest cross-reactivity in the HP BAL Ag test. Classifying these 3 patients as HP BAL false-positive shows a positive agreement of 90.9% (30/33) between HP BAL and HP U/S. Conclusion 16 patients in the BAL cohorts were Ag-positive for HP and BM, demonstrating the inability to discriminate between these dimorphic fungi through Ag testing. HP/BM CSF Ag testing showed no increase in sensitivity when U/S were tested. Minimal increases in sensitivity were observed in BM/HP BAL versus U/S. However, false-positive results in BAL due to cross-reactivity with GM/BDG cannot be ruled out, suggesting limited utility in the invasive CSF/BAL sources.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- From cross-reactivity to clinical utility: rethinking antigen testing for histoplasmosis and blastomycosis.Journal of clinical microbiology · 2026
- Urine detection of Histoplasma antigens among symptomatic people living with HIV (PLWH) in Mato Grosso do Sul, Brazil: A comparative study of EIA and Lateral Flow Assay.Medical mycology · 2026
- Diagnostic accuracy of bronchoalveolar lavage fluid-based testing for pulmonary cryptococcosis: A systematic review and meta-analysis.Diagnostic microbiology and infectious disease · 2026
- 66 Analysis of Real-world Utilization and Diagnostic Accuracy of Histoplasma Antigen EIA Serum and Urine Testing: A Role for Diagnostic StewardshipAmerican Journal of Clinical Pathology · 2026
- Cryptococcal antigen lateral flow: does testing of lower respiratory tract specimens improve diagnostic accuracy for pulmonary cryptococcosis?Microbiology spectrum · 2026
- Tissue distribution of Histoplasma capsulatum in deceased people living with HIV assessed by minimally invasive tissue sampling: An autopsy-based case series in the Brazilian Amazon.PLoS neglected tropical diseases · 2026