Half-count myocardial imaging on the D-SPECT camera: phantom validation and clinical agreement for perfusion and function.
Sharon Gomez Villegas, Dea Dundara Debeljuh, Daniela Calcaterra, Ivan Pribanic, Lucia Leva, Federico Garrou, Andrea D'Alessio, Roberta Matheoud,
Department of Physics, University of Trieste, Via Valerio 2, 34127 Trieste, Italy. · Medical Physics and Radiation Protection Department, University Hospital Rijeka, Rijeka 51100, Croatia.
Key points
From the paper's conclusion- The study demonstrated the feasibility of a 50% count reduction for myocardial perfusion imaging performed with the D-SPECT system and the acquisition/reconstruction protocol evaluated in this study, preserving diagnostic accuracy while lowering radiation exposure or acquisition time.
Abstract
OBJECTIVES: To evaluate the accuracy of the Cadmium-Zinc-Telluride (CZT) D-SPECT (Spectrum Dynamics, Israel) system in quantifying left ventricle (LV) perfusion and function in 99mTc myocardial perfusion imaging (MPI) at half count statistics. METHODS: Segmental perfusion, extent and volume of a perfusion defect were evaluated in an anthropomorphic phantom simulating normal and pathological myocardial condition at the reference 1 million counts (Mc) and at reduced count statistics (0.8, 0.6, 0.4, 0.25Mc).Summed Stress Scores (SSS), Summed Rest Scores (SRS), Summed Difference Score (SDS), Total Perfusion Deficit (TPD), Ischemic Total Perfusion Deficit (ITPD) and Ejection Fraction were evaluated in 326 MPI patient studies, stratified on the classification of myocardial perfusion and LV ejection fraction (LVEF), at 1 and 0.5Mc statistics. RESULTS: In the phantom study, the count statistics did not affect mean perfusion and perfusion defect extent and volume down to 0.4Mc.In the patient study, high concordance was observed between datasets, with Kw values of 0.846 (SSS), 0.922 (SRS), 0.676 (SDS), and K values of 0.89 (TPD) and 0.70 (ITPD). Bland Altman analysis showed minimal bias and narrow limits of agreement for LVEF (-0.1%; -6.9% to 6.7%). CONCLUSIONS: The study demonstrated the feasibility of a 50% count reduction for myocardial perfusion imaging performed with the D-SPECT system and the acquisition/reconstruction protocol evaluated in this study, preserving diagnostic accuracy while lowering radiation exposure or acquisition time. ADVANCES IN KNOWLEDGE: The ability to maintain diagnostic accuracy with 50% fewer counts supports the adoption of optimized, low-dose SPECT protocols enhancing patient safety and imaging efficiency. This reinforces the role of CZT technology in low-dose nuclear cardiology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative and semi-quantitative comparison of myocardial blood flow between 99mTc-tetrofosmin and 99mTc-sestamibi: based on continuous rapid dynamic imaging technology using cadmium-zinc-telluride cardiac single-photon emission computed tomography.Quantitative imaging in medicine and surgery · 2026
- From static to dynamic SPECT/CT: A systematic review of clinical applications and perspectives.Seminars in nuclear medicine · 2026
- Myocardial blood flow quantification by widely available conventional nai SPECT improves detection of hemodynamically significant coronary stenosisEuropean Heart Journal · 2026
- Platform-specific impact of CT attenuation correction on CZT myocardial perfusion SPECT: per-vessel detection of obstructive coronary artery disease.The international journal of cardiovascular imaging · 2026
- Incremental value of dedicated [201Tl] cadmium-zinc-telluride thallium SPECT over rest/stress [99mTc] technetium sestamibi SPECT MPI: a comparative analysis of fixed perfusion defects.Frontiers in cardiovascular medicine · 2026
- Performance of SPECT myocardial perfusion imaging in fulminant myocarditis: comparison to cardiac magnetic resonance imaging.American journal of nuclear medicine and molecular imaging · 2026