Evaluation of the additional occupational risk for workers with orthopedic implants in MRI settings.
Luca Zilberti, Alessandro Arduino, Simone Busoni, Vincenzo Camisa, Vittorio Cannatà, Andrea Magistrelli, Vittorio Miele, Nadia Oberhofer,
Istituto Nazionale di Ricerca Metrologica, Divisione Metrologia dei materiali innovativi e scienze della vita, - Settore Campi e sistemi elettromagnetici, Strada delle Cacce 91, Torino 10135, Italy. · Health Physics Department, Azienda Ospedaliero-Universitaria Careggi, Largo Brambilla 3, Firenze 50134, Italy.
Key points
From the paper's conclusion- As long as the body of the worker with passive non-ferromagnetic orthopedic implants remains outside the MRI scanner, the incremental risk posed by the presence of the implants is negligible. The Lenz effect is weak even if the part of the worker's body with an implant enters the scanner bore.
Abstract
OBJECTIVE: To provide quantitative information that can aid the relevant authority in deciding whether workers with metallic nonferromagnetic orthopedic implants can work in proximity to magnetic resonance imaging (MRI) scanners, including when the exam is being performed. METHODS: Virtual experiments in which the exposure of workers with and without prostheses is evaluated through numerical simulation of realistic human models, with reference to 4 physical effects: radiofrequency (RF)-induced heating, gradient-induced heating, gradient-induced peripheral nerve stimulation, and mechanical actions related to the Lenz effect. RESULTS: In the presence of the considered implants, the extent of RF-induced heating and of the electric field responsible for peripheral nerve stimulation is low and comparable to what is observed without the prostheses. The extent of gradient-induced heating and Lenz effect, which occur only in the presence of implants, is minimal. CONCLUSION: As long as the body of the worker with passive non-ferromagnetic orthopedic implants remains outside the MRI scanner, the incremental risk posed by the presence of the implants is negligible. The Lenz effect is weak even if the part of the worker's body with an implant enters the scanner bore.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards the experimental assessment of exposure to extremely low frequency magnetic field.Physics in medicine and biology · 2026
- RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach.Physics in medicine and biology · 2026
- Head bandaging for patients with auditory implants undergoing MRI, highlighting the risk of magnetic puck projectile.Cochlear implants international · 2026
- Novel method of measuring magnetically induced ForcesCurrent Directions in Biomedical Engineering · 2026
- MRI and Implant Safety at Low‐Field and Ultralow‐Field StrengthsJournal of Magnetic Resonance Imaging · 2025
- Mid-field MRI with cochlear implants: a healthy volunteer study of magnet displacement, pressure perception, and temporal bone imagingBMC Medical Imaging · 2026