Time-resolved sampling of isocyanate peak exposure—chamber validation of a prototype sampler
Daniel Karlsson, Mikael Adamsson, Monika Hassgård
Systems Analytics (United States)
Key points
From the paper's conclusion- The prototype TR-sampler was successfully validated for monitoring of high TDI exposure peaks (≈2 mg/m3 [280 ppb v/v] for a 1-min sample) as well as monitoring of concentrations down to 0.25 µg TDI/m3 [0.036 ppb] for a 1-min sample with an accuracy of > 90% compared to the reference sampler.
Abstract
A prototype sampler for time-resolved monitoring of toluene diisocyanate (TDI) exposure peaks was developed. The sampler contained multiple sampling tubes arranged in a manifold equipped with an internal valve and a step motor, enabling automatic switching of the sampling flow between the tubes, allowing a sequence of up to 10 samples to be collected automatically. The inlet part of the sampling tubes contained a glass wool plug, impregnated with equimolar amounts of di-n-butyl amine (DBA) and acetic acid, to ensure efficient collection and derivatisation of the diisocyanates. The diisocyanate-DBA derivatives were analysed using liquid chromatography tandem mass spectrometry. A streamlined workup procedure was introduced, consisting of a single extraction step combined with the addition of a deuterium-labelled internal standard. The streamlined workup procedure was validated against a reference method (ISO 17734). The sampler tubes were validated by monitoring concentrations from a TDI standard atmosphere and through comparative measurements with a reference method. During short-term sampling (<5 min) at flow rates of up to 2.0 L/min, the sampler tubes exhibited minimal breakthrough and high precision, with a relative standard deviation below 5%. At lower flow rates (0.5 L/min), the tubes were suitable for long-term sampling of up to 4 h. The prototype TR-sampler was successfully validated for monitoring of high TDI exposure peaks (≈2 mg/m3 [280 ppb v/v] for a 1-min sample) as well as monitoring of concentrations down to 0.25 µg TDI/m3 [0.036 ppb] for a 1-min sample with an accuracy of > 90% compared to the reference sampler.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Time-resolved sampling of toluene diisocyanate—peak exposure and long-term monitoring field applicationsAnnals of Work Exposures and Health · 2026
- Geometry-optimized passive SPME sampler for time-weighted average determination of volatile organic compounds in air: Proof of concept.Talanta · 2026
- Analysis of volatile organic compounds by thermal desorption chromatography mass spectrometry (TD-GC-MS): Breakthrough phenomenon, a parameter to take into account.Journal of chromatography. A · 2026
- A respirable cyclone sampler for a flow rate of 20 l/minJournal of Aerosol Science · 2026
- Advantages and limitations of a programmable sampling unit combined with PDMS passive samplers and TD-GC-MS for detection of amphetamine production waste discharges in the sewage system.Forensic science international · 2026
- Task-based transient volatile organic compounds exposure during tanker loading and unloading among hazardous chemical transport workers in South Korea.Annals of work exposures and health · 2026