Thigh abscess following alternate-site continuous glucose monitoring sensor placement.
Kristy M Shaeer, Raechel T White, Benjamin Lord, Tina L Pham, Sarah Kimble, Erin L Reynolds, Jennifer Teeters, Kevin Cowart
University of South Florida Taneja College of Pharmacy, Tampa, FL, and University of South Florida Morsani College of Medicine, Tampa, FL, USA. · University of South Florida Taneja College of Pharmacy, Tampa, FL, USA.
Key points
From the paper's conclusion- This case underscores the potential for a skin and soft tissue infection following off-label CGM placement, especially in patients with uncontrolled hyperglycemia. Pharmacists play a critical role in reinforcing appropriate device use, identifying early signs of infection, and coordinating interdisciplinary care to mitigate device-related complications.
Abstract
PURPOSE: To describe a methicillin-susceptible Staphylococcus aureus (MSSA) thigh abscess following alternate-site placement of a continuous glucose monitor (CGM) sensor and to highlight safety considerations associated with off-label device use in patients with poorly controlled diabetes. SUMMARY: A 49-year-old man with uncontrolled type 2 diabetes (glycated hemoglobin, 11.3%) was referred to an ambulatory care pharmacist for diabetes management. The patient placed the sensor on his upper thigh due to prior dislodgement from the upper arm. Three days later, the patient presented with a painful, erythematous, fluctuant 4.5 x 5 cm abscess at the sensor insertion site. Initial incision and drainage (I&D) was performed and oral sulfamethoxazole/trimethoprim was prescribed. Four days later, he returned with worsening pain and impaired ambulation requiring repeat I&D. Wound cultures grew MSSA. Management additionally included wound packing, topical mupirocin, and continuation of sulfamethoxazole/trimethoprim, with clinical improvement noted after the second I&D procedure. Cutaneous complications with CGM use are typically mild, infrequent, and most often involve contact or irritant dermatitis rather than infection. Reports of CGM-associated skin infections are rare, and none have been reported following sensor placement at unapproved insertion sites. CONCLUSION: This case underscores the potential for a skin and soft tissue infection following off-label CGM placement, especially in patients with uncontrolled hyperglycemia. Pharmacists play a critical role in reinforcing appropriate device use, identifying early signs of infection, and coordinating interdisciplinary care to mitigate device-related complications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Injection-Site Abscess Caused by Mycobacterium abscessus Following Testosterone Injection: A Case ReportCureus · 2026
- Mechanical Shunt Catheter Injury From Repeated Insulin Injections in a Diabetic Patient.Korean journal of neurotrauma · 2026
- Skin and Soft Tissue Infections in the Diabetic Patient.Infectious disease clinics of North America · 2026
- Infusion-Site MSSA as the Source of Confirmed Toxic Shock Syndrome in a 9-Year-Old on Automated Insulin Delivery System: A Case Report.Case reports in pediatrics · 2026
- Intraoperative Continuous Glucose Monitor Sensor-Receiver Connectivity Is Maintained With Alternative Sensor Placement: A Pilot Study.Diabetes, obesity & metabolism · 2026
- Mini-ID - A minimally invasive drainage strategy for complicated skin and soft tissue infection: A case report.Medicine · 2026