American Journal of Clinical Pathology

73 Clinical Utility of Recently FDA-Approved QScoutTM Point-of-care Analyzer for Rapid Screening of Sepsis Patient: Comparison With Other Biomarkers and von Willebrand Factor/ADAMTS13 Ratio

Molly Walkenhorst, Nima Sarani, Keenan Hogan, John Kummer, Raymond Che, Lauren Rowan, Hiral Bhatt, Sarah Gralnek,

Key points

From the paper's conclusion
  • We also observed significantly higher plasma von Willebrand antigen (vWF) in the sepsis group compared to the non-disease group (mean: 40.2 mg/mL vs 10.7 mg/mL group).

Abstract

Background Sepsis is a life-threatening disease where quick diagnosis saves life. Recently FDA approved QScoutTM analyzer (Ad Astra Diagnostics, Morrisville, NC) which utilizes a cutting-edge imaging-based system where cells are examined directly by an AI-trained system instead of indirect measurements. QScout™ is the first point-of-care analyzer to report immature granulocytes and neutrophil-to-lymphocyte ratio (NLR), that can provide insight on infection status and severity. We investigated clinical utility of this test using 44 patients suspected of sepsis (18 sepsis group, mean age 68.6 years, 10 males and 8 females and 26 non-disease group, mean age 49.0 years, 12 males, 14 females). Materials and Methods QScoutTM can analyze EDTA whole blood within 10 minutes providing various CBC parameters including calculated percentage immature granulocytes (%IG) and NLR. Left over plasma specimens from chemistry lab were used for measuring plasma von Willebrand factor (vWF) antigen and ADAMTS13 antigen by ELISA assays. Monocyte distribution width (MDW) values were obtained during routine CBC analysis using Beckman Coulter analyzer. Results MDW (28.1% higher), WBC count (118.2% higher), %IG (263.6% higher) and NLR(332.2% higher) were significantly elevated in sepsis group compared to non-disease group by both independent t-test, two tailed (p < 0.05 in all groups) and Mann-Whitney U test two tailed (p < 0.05 in all groups). For example, reference range of % IG is up to 0.7%, but in sepsis group, %IG mean was 1.2 compared to 0.33 in non-disease group. Similarly, reference range of NLR is 0.86-3.83, but in sepsis group the mean NLR was 14.7 compared to mean value of 3.4 in non-disease group. Moreover, recently developed Complete Blood Count Sepsis Index (CBC-SI) was zero in all but one in non-disease group (mean: 0.08). In contrast, CBC-SI varied from 1 to 5 in 17 out of 18 patients in sepsis group (mean:2.2) . We also observed significantly higher plasma von Willebrand antigen (vWF) in the sepsis group compared to the non-disease group (mean: 40.2 mg/mL vs 10.7 mg/mL group). In contrast, percentages of ADAMTS13 activity were lower in sepsis group than non-disease group (67.8% vs 81.23%). However, vWF/ADAMTS13 ratio was significantly higher in the sepsis group compared to non-disease group (mean: 7.2 in sepsis group vs 1.4 in non-disease) and difference was significant by both t-test and Mann-Whitney U test (p < 0.05) Discussion QScoutTM provided fast CBC result including calculated NLR, and %IG which are helpful in screening patients suspected of sepsis at the emergency room. Moreover, CBC-SI index calculated using WBC, NLR and MDW is also useful in evaluating sepsis. Significant increase in plasma vWF in sepsis patients indicate that vWF may be a surrogate marker for endotheliopathy in sepsis. Moreover, significantly higher vWF/ADAMTS13 in disease group indicated potential risk of microvascular thrombosis in sepsis patients.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers