Comparisons of seismocardiographic sensor positions for VO2max estimations.
Sofie Braun Svensen, Julie Kathrine Dittmann Kathrine Dittmann Andreasen, Joan Hentze, Emil Korsgaard, Rasmus Gundorff Sæderup, Peter Søgaard, Ahmad Agam, Samuel Emil Schmidt
Medicine, Aalborg University Department of Health Science and Technology, Selma Lagerløfs Vej 249, 9270 Gistrup, Aalborg, North Denmark Region, 9260, Denmark. · Department of Health Science and Technology, Aalborg University, Fredrik Bajers Vej 7K, Aalborg, Region Nordjylland, 9220, Denmark.
Key points
From the paper's conclusion- VO 2 max estimates showed the strongest agreement with the XP reference at the right thorax, left sternum, and lower left thorax overlying the cardiac silhouette compared with the remaining measurement locations. 

These findings suggest that several alternative SCG sensor locations provide VO 2 max estimates comparable to the standard XP location, whereas a limited number of locations pro
Abstract
VO2max is a key health indicator of cardiorespiratory fitness and overall physical performance. VO2max can be estimated using seismocardiography (SCG). SCG can be measured at multiple anatomical locations. The xiphoid process (XP) has primarily been used for VO2max estimation, and the comparative performance of alternative sensor locations has not been evaluated. The aim of this study was to determine whether SCG sensor locations other than the XP produce systematically different VO2max estimates when the same estimation algorithm is applied. The study included 13 healthy male participants. SCG signals were sampled from a 4 × 4 grid of accelerometers, resulting in 16 different sensor locations across the chest, including the XP. VO2max was estimated using a previously validated non-exercise SCG algorithm based on fiducial SCG features and demographic variables. Differences in estimated VO2max between each sensor location and the XP reference were analysed using a linear mixed-effects model, and Pearson's correlation coefficients were calculated to assess the relationship between sample locations. The analysis showed that seven sensor locations differed significantly from the XP reference, with mean deviations ranging from 1.108 to 2.626 ml·kg⁻¹·min⁻¹. All correlations were greater than 0.849, indicating high consistency between estimates. Bland-Altman analyses demonstrated small mean biases ranging from -2.47 to 2.03 ml·kg⁻¹·min⁻¹ and no evidence of proportional bias for any sensor location, although the limits of agreement indicated some variability at the individual level. VO2max estimates showed the strongest agreement with the XP reference at the right thorax, left sternum, and lower left thorax overlying the cardiac silhouette compared with the remaining measurement locations. These findings suggest that several alternative SCG sensor locations provide VO2max estimates comparable to the standard XP location, whereas a limited number of locations produce systematic differences and may require location-specific correction before clinical implementation.
.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of cardiopulmonary function in patients with scoliosis at rest and during exercise: a case-control study.Frontiers in surgery · 2026
- CaReFit: Cost-Effective Cardiorespiratory Fitness Assessment Using Multimodal SensorsProceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies · 2026
- Exploring seismocardiographic signals in children with congenital heart disease: a prospective pilot studyScientific Reports · 2026
- Sensor Technologies for Wearable and Portable Cardiac Acoustic Monitoring: An Experimental Study.Medical engineering & physics · 2026
- Noninvasive tracking of right ventricular systolic pressure via multi-channel digital auscultationEuropean Heart Journal · 2026
- reSPIRE: A wearable multimodal sensing system for monitoring ventilation, cardiovascular dynamics, and respiratory muscle activityBiosensors and Bioelectronics · 2026