Comparison of artifact elimination strategies during postprocessing for the numerical assessment of localized low frequency magnetic field exposure of the hand.
Gernot Schmid, Pia Schneeweiss, Rene Hirtl
EMC & Photonics, Seibersdorf Laboratories, Tech Campus Seibersdorf, Seibersdorf, Lower Austria, 2444, Austria. · EMC & Photonics, Seibersdorf Laboratories, Tech Campus Seibersdorf, Seibersdorf, 2444, Austria.
Key points
From the paper's conclusion- Our results clearly demonstrate the inappropriateness of the 99th percentile-based assessment, as it may lead to underestimations of the actual induced E i,avg by more than a factor of 100, depending on the exposure scenario.
Abstract
Numerical computations of induced electric field strength Eiinside the human body are prone to so-called "staircasing errors", i.e., physically incorrect local enhancements of the computed Eiate such numerical artifacts, ICNIRP suggested in their 2010 guidelines to use the 99thpercentile of the distribution of the 2 x 2 x 2 mm3at the edges of the cubical cells. Similar artifacts can occur due to imperfections in the anatomical body models used. In order to elimin averaged induced electric field strength Ei,avg for comparison with the basic restrictions instead of the maximum computed Ei,avgvalue. However, this suggestion was based on very early computational studies using anatomical body models with relatively low spatial resolution. Moreover, a 99thpercentile-based assessment of Ei,avgappears generally questionable in case of highly localized exposure scenarios, where large fractions of the body remain almost unexposed, and therefore true local hot spots of Ei,avg that should be considered may be excluded from the analysis, simply because the exposed tissue fraction is less than 1 % of the total body volume. Although this problem has been repeatedly pointed out by several research groups in the past, the 99thpercentile-based assessment unfortunately spread out as a "general rule" into some guidelines for exposure assessment. Consequently, the compliance of many devices on the market with exposure limits can be questioned. In this paper we are quantifying the potential underestimation of exposure when using the 99thpercentile-based Ei,avg for several representative scenarios of a highly localized magnetic field exposure of the hand using a state-of-the-art anatomical body model. Besides comparing the 99thpercentile-based Ei,avg values with values obtained when using higher percentiles, we also compare with Ei,avgvalues obtained by applying a statistics-based outlier elimination method proposed by Arduino et al. (2020). Our results clearly demonstrate the inappropriateness of the 99th percentile-based assessment, as it may lead to underestimations of the actual induced Ei,avgby more than a factor of 100, depending on the exposure scenario. In contrast, the statistics-based artifact elimination strategy proves to be a reasonable and robust method for estimating the actual value Ei,avg for highly localized magnetic field exposure of the hand.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Towards the experimental assessment of exposure to extremely low frequency magnetic field.Physics in medicine and biology · 2026
- RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach.Physics in medicine and biology · 2026
- Meta-topological hydrogel enables multisource and frequency-tailored artefact mitigation for bioelectronicsNature Sensors · 2026
- Spatiotemporal Modeling of Impedance Degradation and Compensation of ERT Electrodes for Target Reconstruction.Sensors (Basel, Switzerland) · 2026
- Gaussian decomposition of EIT reconstructions with different inclusion geometries.Physiological measurement · 2026
- Electrocardiographic Imaging via Posterior Sampling under Operator DiscrepancyarXiv · 2026