On the feasibility of temporal interference stimulation of human brains using two arrays of electrodes.
Yu Huang
Research and Development, Soterix Medical Inc, 1480 US Highway 9 North, #204, Woodbridge Township, New Jersey, 07095, United States.
Key points
From the paper's conclusion- We restate the focality-intensity tradeoff in the context of TI and provide a first voxel-level map (at 4 mm resolution) of achievable TI focality and modulation strength in the MNI-152 head template.
Conclusions: Compared to two-pair or multi-pair TI, we promote two-array TI for its similar performance in focality and lower cost in both optimization time and electrodes needed.
Abstract
Temporal interference (TI) stimulation leverages two pairs of electrodes to induce an interfering electrical field in the brain. Studies show that TI can stimulate the deep brain without affecting shallow areas. Optimal placement of these electrodes for maximal focal modulation remains computationally challenging. We are the first to propose two arrays instead of two or multiple pairs of electrodes to boost modulation focality. However, it outputs too many electrodes with overlaps across two frequencies, making it difficult to implement in practice. Objective: Based on recent progress in developing multi-channel TI devices and computational work, here we advocate two-array TI with feasibility data. Methods & Results: We review TI optimization algorithms, and compare them over 25 heads across six brain targets. We show that the latest optimization algorithm for two-pair TI innately works for two-array TI with the fastest speed (under 30s) and a similar amount of electrodes as in multi-pair TI. At four of the six targets, this fastest algorithm for two-array TI achieves similar or even better focality than TI using 16 pairs of electrodes that takes days to optimize. We show a hardware implementation of two-array TI on our 8-channel TI device. We argue that two-pair TI is only preferred for maximal modulation strength or when hardware is limited to two current sources. We restate the focality-intensity tradeoff in the context of TI and provide a first voxel-level map (at 4 mm resolution) of achievable TI focality and modulation strength in the MNI-152 head template. Conclusions: Compared to two-pair or multi-pair TI, we promote two-array TI for its similar performance in focality and lower cost in both optimization time and electrodes needed. We hope this work will pave the way for future adoptions of two-array TI for more focal non-invasive deep brain stimulation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Replacing continuous stimulation of one set of electrodes with successive stimulation of multiple sets of electrodes can improve the focality of transcranial temporal interference stimulation (tTIS), especially the focality of stimulation towards deep brain regions: A simulation study.bioRxiv · 2026
- Suprathreshold suppression underlies spatial selectivity in temporal interference stimulationbioRxiv · 2026
- Non-invasive, high-precision, and independently controllable neuromodulation via optimized multiple temporal interference stimulationScience Advances · 2026
- Artefact-free EEG during temporal interference stimulation reveals phase-specific modulation of the human visual cortexbioRxiv · 2026
- Field-Based Characterization of Temporal Interference Stimulation: Beyond the Target-Centric PerspectivebioRxiv · 2026
- Focal and Steerable Temporal Interference Stimulation of the Human Left Posterior Insula: A Proof-of-Concept Validation Using Intracranial Recordings in a Human Cadaver.Bioengineering (Basel, Switzerland) · 2026