Discrepancy between Collapsed Cone and Monte Carlo algorithms in lung inhomogeneity correction.
Amir Hossein Karimi, Indra J Das, Saeedeh Moharramkhani, Hossein Mirzaei, Marco Esposito, Ghazale Geraily
Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences School of Medicine, Tehran, Tehran, 0, Iran (The Islamic Republic of). · Department of Radiation Oncology, Northwestern Medicine, 231 Huron St, Chicago, Illinois, 60611, United States.
Key points
From the paper's conclusion- The aim of this study is to evaluate the accuracy of Collapsed Cone (CC) algorithm for dose calculation in the presence of lung inhomogeneity compared to the Monte Carlo (MC) algorithm as the reference.
Abstract
Stereotactic body radiation therapy (SBRT) is a well-established standard of treatment for non-small cell lung cancer patients. However, SBRT is a double-edged sword and a high degree of accuracy is required for dose calculations. In the presence of inhomogeneity, it should be below 3% and depends strongly on the accuracy of the algorithm used for dose calculation in treatment planning system (TPS). The aim of this study is to evaluate the accuracy of Collapsed Cone (CC) algorithm for dose calculation in the presence of lung inhomogeneity compared to the Monte Carlo (MC) algorithm as the reference. The study conducted on a wide range of photon fields from 0.5×0.5 cm2 to 10×10 cm2. Monaco® TPS equipped with CC and MC algorithms in 6 MV, 10 MV, and 18 MV photons was employed and the inhomogeneity correction factor (ICF) and PPDs were calculated inside an in-house inhomogeneous phantom. Finally, the results were compared between two algorithms. For 6 MV photons, gamma index pass rate (GIPR) with (2 mm, 3%) criteria in a 0.5×0.5 cm² field is only 71.3%. The CC algorithm won't be able to confidently reproduce the PDDs in this field. For 1×1 cm², 2×2 cm², and 3×3 cm² fields, the accuracy of the CC algorithm can increase to above 98%. For SBRT of small lung lesions, the results recommend increasing the minimum segment width to > 0.5 cm in IMRT plans if CC algorithm is used. Independent of field size, the accuracy of CC algorithm decreases with increasing energy. For 10 MV and 18 MV photons, average GIPR under (2 mm, 3%) criteria was 97.9% and 92.0%, respectively, which are not of clinical interest.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of grid size and statistical uncertainty parameters on dose calculations using the Monte Carlo algorithm in the Monaco treatment planning system for liver stereotactic body radiotherapy.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- How robust are secondary dose algorithms for Patient-Specific quality assurance in modulated stereotactic Radiosurgery?Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Impact of Volume and Image Strategy on Dose Prescription Accuracy in Lung Stereotactic Body Radiation Therapy.Practical radiation oncology · 2026
- Influence of Delivery Geometry and Beam Modulation on Dose Gradient and Organ-at-Risk Sparing in Lung SBRT: A Comparative Planning StudyPolish Journal of Medical Physics and Engineering · 2026
- Dosimetric comparison of the Anisotropic Analytical Algorithm and Acuros XB for single-fraction cranial stereotactic radiosurgery with a 6 MV flattening filter-free beam.Biomedical physics & engineering express · 2026
- Comparing Strategies for Pencil Beam Scanning Carbon Ion Radiation Therapy in Stage I Peripheral Non-Small Cell Lung Cancer: Toward Optimized Clinical Decision-Making Planning.International journal of radiation oncology, biology, physics · 2026