Single-scan structured-illumination dual-modal method for high-completeness micro/nano surface topography measurement
Ling Hu, Jinze Li, Linlin Zhu, Xinghua Li
Key points
From the paper's conclusion- Six independent measurements of a step-height standard with a nominal height of 1.0 μ m yielded a standard deviation of 0.0299 μ m in the measured step heights.
Abstract
Complex micro/nano surfaces can produce locally unreliable responses and missing height data in optical measurements. This paper presents a single-scan structured-illumination dual-modal microscopy method that improves reconstruction completeness by reusing the same acquired images for two complementary reconstruction pathways. At each axial position, two phase-shifted fringe images are acquired. Two-frame differencing and one-dimensional Hilbert transform yield a computational optical-sectioning response for height reconstruction, termed CM reconstruction in this work. The first image acquired at each axial position is also used to construct an active focus variation (FV) response from local fringe carrier-band energy. No separate FV image sequence or additional spatial registration is required. The contrast-to-noise ratio of each response is used to assess reliability and guide height selection and weighting. Experiments on micro-bumps of three sizes, grinding wheel abrasive grains, and micro imprinted deep holes showed that fusion increased valid data coverage compared with either reconstruction pathway alone in all tested fields of view. Local comparisons showed that FV supplied heights where CM sidewall data were missing, while fusion retained local features from reliable CM results. Six independent measurements of a step-height standard with a nominal height of 1.0 μ m yielded a standard deviation of 0.0299 μ m in the measured step heights.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Continuous 3D microscopic measurement via spatial-resolution compensation.Optics letters · 2026
- All-telecentric multi-camera microscopic FPP: Scheimpflug calibration and reprojection-based multi-view fusionOptics Express · 2026
- A Robust Method for Microscopic 3D Shape Restoration via Shape-From-Focus.Microscopy research and technique · 2026
- Depth-synchronized adaptive coherence scanning interferometry for rapid 3D morphology of composite high-aspect-ratio micro-trenches.Optics letters · 2026
- Unified variational framework for simultaneous correction of non-uniform sampling and diffraction-induced slope errors in coherence scanning interferometry.Journal of the Optical Society of America. A, Optics, image science, and vision · 2026
- Polarization-constrained ring-illumination microscopy for label-free relative 3D surface reconstruction.Optics letters · 2026