Medical engineering & physics

Rhythm priors and beat-specific cardiomechanical information in ECG-anchored PCG-envelope estimation.

Yingwei Li, Qianxiang Zhou, Weiwei Ding, Zhongqi Liu

Beihang University, 37 Xueyuan Road, Haidian District, Beijing, 100091, China. · Beihang University, 37 Xueyuan Road, Haidian District, Beijing, Beijing, 100091, China.

Key points

From the paper's conclusion
  • Electrocardiographic (ECG) R alignment and cardiac rhythm can supply strong priors for phonocardiographic (PCG) envelope estimation, complicating interpretation of accuracy as beat-specific mechanical information.
Methods: Frozen participant-held-out accelerometer (ACC) predictions covered 12,000 beats from 40 healthy participants.

Abstract

Electrocardiographic (ECG) R alignment and cardiac rhythm can supply strong priors for phonocardiographic (PCG) envelope estimation, complicating interpretation of accuracy as beat-specific mechanical information. Methods: Frozen participant-held-out accelerometer (ACC) predictions covered 12,000 beats from 40 healthy participants. Experiment A reassigned predictions across 100 same-participant, same-fold wrong-beat mappings approximately matched on neighboring RR intervals (ECG R-to-R intervals). Experiment B used training-only nested-selected continuous RR templates. The three principal outcomes were envelope mean absolute error (MAE), correlation, and corrected second-heart-sound (S2) timing MAE. Corrected S2 evaluation excluded seven references outside the fixed 180-520-ms window (11,993 eligible events), distinct from label correction or multiplicity adjustment. Benjamini-Hochberg correction covered all six original comparisons, with unchanged prespecified stability rules. Results: True correspondence improved MAE by 0.0096 (95% CI 0.0080-0.0111; 40/40 participants; q=7.28×10⁻¹²). ACC improved MAE over continuous RR by 0.0093 (95% CI 0.0036-0.0150; 29/40; q=0.00252). Only MAE met complete stability under both controls. Correlation and corrected S2 showed favorable average effects with lower participant consistency in Experiment B; first-heart-sound (S1) timing remained an unfavorable outcome. Additional descriptive sensitivities retained average MAE benefits under expanded bandwidths (0.0085; 28/40) and balanced dual-RR matching (0.0097; 40/40). The expanded grid reduced the original B advantage by 8.262% without eliminating the average ACC benefit. Conclusions: Complementary rhythm and correspondence controls support an outcome-specific envelope-MAE increment associated with the true mechanical beat. The evidence concerns offline, non-causal, ECG R-anchored normalized-envelope estimation in one healthy cohort; clinical and external validation remain future requirements.

Read the paper