Design and Repeatability Testing of a Novel Benchtop Apparatus for Evaluating Wear Durability of Artificial Tendons.
Shane Miller Abercrombie, Pierre-Yves Mulon, David Edgar Anderson, Dustin Lee Crouch
Biomedical Engineering, University of Tennessee Knoxville Tickle College of Engineering, 1506 Middle Drive, Knoxville, Tennessee, 37996, United States. · The University of Tennessee Knoxville College of Veterinary Medicine, 2407 River Drive, Knoxville, Tennessee, 37996, United States.
Key points
From the paper's conclusion- Under lubricated conditions, wear scar morphology was consistent with isolated mechanical fiber abrasion, and median UTS was 865.7 N (848.0-887.2 N), with high inter-trial repeatability across three independent tests (p = 0.3688).
Abstract
Artificial tendons are potential alternatives to biological grafts for reconstructing large tendon defects, yet their clinical translation is limited partly by mechanical durability during repetitive gliding against pulley-like structures (e.g., retinacula). Standardized methods for evaluating long-term tendon-pulley wear under physiologically relevant conditions are lacking. This study aimed to develop a benchtop testing methodology for mechanical wear of artificial tendons after high-cycle loading. A custom apparatus was designed to simulate key biomechanical features of the flexor pollicis longus tendon, including tensile load, excursion velocity, and wrapping angle at the tendon-pulley interface. Apparatus performance and repeatability were evaluated using polyester tendons subjected to 1.02 million simulated thumb flexions, equivalent to approximately 395 days of simulated physiological thumb motion, at a rate of 22.1 simulated flexions per second. Non-lubricated testing resulted in excessive frictional heating (93 °C within 40,800 simulated thumb flexions), causing premature thermal damage and ultimate tensile strength (UTS) variability. Conversely, applying boundary lubricant at the tendon-pulley interface, mimicking synovial fluid, maintained temperatures at approximately 36 °C and eliminated non-physiological thermal degradation. Under lubricated conditions, wear scar morphology was consistent with isolated mechanical fiber abrasion, and median UTS was 865.7 N (848.0-887.2 N), with high inter-trial repeatability across three independent tests (p = 0.3688). This study establishes a reproducible experimental framework for isolating mechanical wear in artificial tendons under high-cycle gliding and is suited for orthopaedic research applications, including material screening and design optimization of tendon grafts and tendon-implant interfaces.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Early-stage calcification promotes tendon damage under cyclic loading through alterations in interfascicular matrix viscoelastic properties.Journal of the mechanical behavior of biomedical materials · 2026
- Quantifying the role of hyaluronic acid on collagen fiber mechanics via femtosecond laser ablation.bioRxiv : the preprint server for biology · 2026
- Modeling tendon viscoelastic properties using a system identification approach.Journal of biomechanics · 2026
- Evaluation of a novel strap fixation method for biomechanical testing of sheep Achilles tendons.Archives of orthopaedic and trauma surgery · 2026
- Impact of Test Speed and Lubrication Conditions on Dynamic Testing of Total Knee EndoprosthesesLubricants · 2026
- Fatigue Resistance of Cardiac Lead Insulations - A Novel in vitro Test Bench Mimicking Physiological LoadingCurrent Directions in Biomedical Engineering · 2026