Medical engineering & physics

Design and Repeatability Testing of a Novel Benchtop Apparatus for Evaluating Wear Durability of Artificial Tendons.

Shane Miller Abercrombie, Pierre-Yves Mulon, David Edgar Anderson, Dustin Lee Crouch

Biomedical Engineering, University of Tennessee Knoxville Tickle College of Engineering, 1506 Middle Drive, Knoxville, Tennessee, 37996, United States. · The University of Tennessee Knoxville College of Veterinary Medicine, 2407 River Drive, Knoxville, Tennessee, 37996, United States.

Key points

From the paper's conclusion
  • Under lubricated conditions, wear scar morphology was consistent with isolated mechanical fiber abrasion, and median UTS was 865.7 N (848.0-887.2 N), with high inter-trial repeatability across three independent tests (p = 0.3688).

Abstract

Artificial tendons are potential alternatives to biological grafts for reconstructing large tendon defects, yet their clinical translation is limited partly by mechanical durability during repetitive gliding against pulley-like structures (e.g., retinacula). Standardized methods for evaluating long-term tendon-pulley wear under physiologically relevant conditions are lacking. This study aimed to develop a benchtop testing methodology for mechanical wear of artificial tendons after high-cycle loading. A custom apparatus was designed to simulate key biomechanical features of the flexor pollicis longus tendon, including tensile load, excursion velocity, and wrapping angle at the tendon-pulley interface. Apparatus performance and repeatability were evaluated using polyester tendons subjected to 1.02 million simulated thumb flexions, equivalent to approximately 395 days of simulated physiological thumb motion, at a rate of 22.1 simulated flexions per second. Non-lubricated testing resulted in excessive frictional heating (93 °C within 40,800 simulated thumb flexions), causing premature thermal damage and ultimate tensile strength (UTS) variability. Conversely, applying boundary lubricant at the tendon-pulley interface, mimicking synovial fluid, maintained temperatures at approximately 36 °C and eliminated non-physiological thermal degradation. Under lubricated conditions, wear scar morphology was consistent with isolated mechanical fiber abrasion, and median UTS was 865.7 N (848.0-887.2 N), with high inter-trial repeatability across three independent tests (p = 0.3688). This study establishes a reproducible experimental framework for isolating mechanical wear in artificial tendons under high-cycle gliding and is suited for orthopaedic research applications, including material screening and design optimization of tendon grafts and tendon-implant interfaces.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers