HearTec: An Embedded Dual-Site Oscillometric Platform for Blood Pressure and Arterial Stiffness Monitoring - Design and Preliminary Feasibility Study.
Tahar Omari, Ismael Baji, Florine Caulier, Mehdi Rochd, Stephane Carlier
Electronics and Telecommunications, University of Ain Temouchent Belhadj Bouchaib, Ain Témouchent, Ain Temouchent, Aïn Témouchent Province, 46000, Algeria. · Université Libre de Bruxelles, Bruxelles, Brussels, 1050, Belgium.
Key points
From the paper's conclusion- A key
hardware contribution is the implementation of a dual-solenoid-valve pneumatic architecture
enabling active stabilization of cuff pressure at the mean arterial pressure (MAP), which
provides a stable window for reliable waveform acquisition and pulse transit time analysis.
The embedded platform, based on an Arduino microcontroller, performs pneumatic control,
conditioning
Abstract
Arterial stiffness is an independent predictor of cardiovascular risk, yet traditional blood pressure monitors are inherently unable to assess it, since they perform only single-site, single-cuff measurements and do not capture the pulse wave propagation dynamics required for its estimation. Existing systems capable of arterial stiffness assessment remain expensive, bulky, and confined to specialised clinical settings, limiting their accessibility for routine or personal use. This work presents HearTec, a compact and low-cost biomedical instrumentation platform for dual-site oscillometric acquisition at the brachial and ankle locations, designed to be affordable enough for personal use outside the clinical environment. The system integrates synchronized pneumatic modules, piezoresistive pressure sensors, and dedicated analog front end circuits designed to separate DC cuff pressure from AC oscillometric pulsations. A key hardware contribution is the implementation of a dual-solenoid-valve pneumatic architecture enabling active stabilization of cuff pressure at the mean arterial pressure (MAP), which provides a stable window for reliable waveform acquisition and pulse transit time analysis. The embedded platform, based on an Arduino microcontroller, performs pneumatic control, conditioning circuits, and wireless data transmission via Bluetooth with simultaneous local storage. These preliminary results demonstrate the technical feasibility of signal acquisition only; they do not establish the clinical or physiological validity of the estimated parameters and do not constitute a clinical validation.
.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A physics-informed peripheral hemodynamic digital twin for personalized estimation of arterial stiffnessnpj Digital Medicine · 2026
- Noninvasive pulse wave velocity measurement: technological progress, clinical validation, and translation challenges.Physiological measurement · 2026
- Feasibility of Accelerometer-Based Arterial Pulse Measurement: Part II-Experimental Investigation.Sensors (Basel, Switzerland) · 2026
- Personalized haemodynamic modelling for non-invasive quantification of arterial stiffness in the human vasculature.Journal of the Royal Society, Interface · 2026
- A conformal piezoelectric microsystem for demographic-adaptive and calibration-free cuffless blood pressure monitoringNature Communications · 2025
- H ygieia : A Wearable Metamaterial Patch for Calibration-free Carotid Hemodynamics MonitoringProceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies · 2026