Patient and Database Shift in Lightweight ECG Arrhythmia Classification: A Controlled Evaluation of Noise, Quantization, and Embedded Inference.
Hyun Mok Park, Woojin Cho, Seok-Oh Bang, Hyun-Seok Choi, Ki-Tae Kwon, Jin-Sung Roh, Sang-Wuk Shin, Jong-Min Lim
Digital Convergence Research Headquarters, KIEL Institute, 487 Sosa-ro, Wonmi-gu, Bucheon-si, Gyeonggi-do, Bucheon, 14560, Korea (the Republic of). · Digital Convergence Research Headquarters, KIEL Institute, 487, Sosa-ro, Wonmi-gu, Bucheon-si, Gyeonggi-do, Bucheon, 14560, Korea (the Republic of).
Key points
From the paper's conclusion- We conducted a controlled benchmark of three convolutional architectures under a corrected subject-disjoint MIT-BIH split, external INCART testing, Noise Stress Test Database artifacts, 16-bit floating-point (FP16) and full-integer 8-bit (INT8) conversion, and classifier-only invocation on an ESP32 without external RAM.
Abstract
Efficient microcontroller execution does not ensure reliable electrocardiogram classification across patients and databases. We conducted a controlled benchmark of three convolutional architectures under a corrected subject-disjoint MIT-BIH split, external INCART testing, Noise Stress Test Database artifacts, 16-bit floating-point (FP16) and full-integer 8-bit (INT8) conversion, and classifier-only invocation on an ESP32 without external RAM. Five fixed seeds were used. On the MIT-BIH test cohort (DS2), INT8 four-class macro-F1 was 0.474-0.498 and supraventricular (S) recall was 0.150-0.180. Mean S-class area under the receiver operating characteristic curve (AUROC) was below 0.5 in all six architecture-regime groups, although all corresponding record-cluster intervals included 0.5. Within-record ranking could remain high despite poor pooled ranking. On INCART, S-class ranking was stronger, with all six patient-cluster AUROC intervals above 0.5, but the representative mixed-noise depthwise-separable network (DS-CNN) achieved 0.797 recall with 0.074 precision. Including record 201, which shares a subject with test record 202, increased DS2 S-recall point estimates by 2.49-14.68 percentage points. In the primary architecture panel, FP16 and INT8 met the engineering tolerances in all 120 comparisons. DS-CNN required 9.49 ms per classifier invocation on the ESP32 using preprocessed, annotation-centered inputs. Acquisition, QRS detection, preprocessing, current, and energy were not measured. Patient-dependent S-score behavior and external false-positive predictions limited this panel more than numerical precision or microcontroller resources.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-Time Cardiac Arrhythmia Classification Using TinyML on Ultra-Low-Cost Microcontrollers: A Feasibility Study for Resource-Constrained EnvironmentsBioengineering · 2026
- Evaluation Protocols and On-Chip Feasibility in Embedded Electrocardiogram Classification: A Systematic ReviewElectronics · 2026
- ArrythML-II: An Optimized TinyML Framework for ECG Arrhythmia Detection and Multi-Class Beat Classification on Wearable DevicesarXiv · 2026
- Hybrid feature engineering for resource-efficient Arrhythmia detection in electrocardiogram signals: An interpretable, separability-driven framework.Computers in biology and medicine · 2026
- A Fast and Generalizable Deep Neural Network for the Detection of Atrial Fibrillation.Bioengineering (Basel, Switzerland) · 2026
- Single-Lead ECG Arrhythmia Classification with Machine Learning for IoT HealthcareApplied System Innovation · 2026