科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Medical engineering & physics2026-08-20

Finite Element Mechanical Analysis of Rigid Internal Fixation Selection for Distal Femur Fractures.

Meng Li, Zicheng Zhang, Ming Chen, Jiantao Li, Wei Zhang

原始摘要(英文原文)· Original abstract
Compared with lateral locking plate (LLP) fixation, dual‑plate (DP) fixation has not demonstrated superior clinical outcomes in the treatment of distal femur fractures (DFFs). We propose that this discrepancy arises from unclear indications. Methods: Finite element analysis was conducted to compare the biomechanical performance of LLP-only versus DP fixation in AO/ASIF type C2 fractures with varying metaphyseal defect sizes (0, 5, 10, 15, 20, 25, and 30 mm). Fourteen models were established and subjected to axial (700 N) and torsional (14 Nm) loading. Results: As the metaphyseal defect size increased (0-30 mm), all measured parameters under axial compression and torsional loading exhibited consistent trends. Overall, changes were more pronounced in the LLP-only group. Under axial loading, increasing defect size led to a 95.2% reduction in axial stiffness, a 58.2-fold increase in axial displacement, and 7.7 times rise in the maximum equivalent stress of the LLP-only fixation. Conclusions: This study demonstrates that larger metaphyseal defects result in decreased fixation stiffness, increased fracture-site displacement, and elevated plate stress, including in the medial plate of DP constructs. These changes are consistently less pronounced with DP fixation compared to LLP -only constructs. Biomechanically, a transitional range exists between the two methods, with a mechanical suitability threshold near a metaphyseal defect size of 15 mm.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Finite Element Mechanical Analysis of Rigid Internal Fixation Selection for Distal Femur Fractures. — 科研速览 Science Skim