Analysis of modulation transfer function in large-field-of-view pulsed radiation imaging systems: component contributions and magnification effects

Dong Li, Weipeng Yan, Yan Song, Keqing Gao, Zhen Liu, Langlang Li, Kaixiang Yang, Yang Li,

Key points

From the paper's conclusion
  • We integrate MTF models of the thick pinhole (ray tracing), scintillator (Geant4 simulation), optical lens, gated image intensifier, fiber optic taper, and camera, and construct a comprehensive system MTF expression.

Abstract

Pulsed radiation imaging systems are essential diagnostics in high-energy-density physics, inertial confinement fusion, Z-pinch experiments, and flash X-ray radiography. These systems typically consist of a cascaded chain: a thick pinhole, scintillator, optical lens, gated image intensifier, fiber optic taper, and camera. The overall spatial resolution is jointly determined by the modulation transfer functions (MTFs) of individual components and their coupling under various magnifications. However, most existing studies focus on single components, and a system-level theoretical framework that integrates the component-level MTFs for such pulsed radiation imaging chains remains lacking. In this work, we propose a system-level analysis framework based on component-level MTF modeling. We integrate MTF models of the thick pinhole (ray tracing), scintillator (Geant4 simulation), optical lens, gated image intensifier, fiber optic taper, and camera, and construct a comprehensive system MTF expression. The effects of key parameters — pinhole magnification, image intensifier size, and the number of camera pixels — on the limiting object-plane resolution are systematically analyzed. Results show that pinhole magnification is the dominant factor: when magnification increases from 0.5× to 2.0×, the limiting object-plane resolution (at 5% contrast) improves from 1.2 lp/mm to 2.63 lp/mm. In contrast, for the configurations studied, increasing the image intensifier diameter from 18 mm to 40 mm yields less than 3% improvement, and increasing the number of camera pixels beyond 2K× 2K gives negligible benefit for the baseline configuration. This work identifies resolution bottlenecks under different configurations and provides a practical theoretical tool for system parameter optimization in pulsed radiation imaging diagnostics.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers