Investigating the behavior of hydrogenated amorphous silicon under UHDR proton beam
K. El Achi, A. Warnier, A. Khalilzadeh, D. Radmanovac, S. Waid, C. Schmitzer, S. Rossomme, V. De Beco,
Key points
From the paper's conclusion- The results were reproducible and linear with dose.
Abstract
Radiation-based therapies are crucial treatments in the fight against malignant cancers. Ultra-high dose rate (UHDR) proton therapy promises a higher radiobiological advantage and increased effectiveness in targeting tumor cells while sparing healthy cells by providing the total dose in a short time at a rate higher than 40 Gy/s. This work aims to characterize a hydrogenated amorphous silicon (a-Si:H) pixel under UHDR proton beams. A-Si:H gains particular interest due to its capacity to produce large-scale and high-resolution flat panel detectors at an affordable cost with high radiation hardness. A test board of a 2.4 × 2.4 mm 2 a-Si:H pixel array was acquired and bonded to test at MedAustron in Wiener Neustadt. The test plan was to irradiate using a single spot of constant dose at different spill lengths and spill numbers to test the behavior from conventional to flash regimes for dose rate dependence measurements. Additionally, an ionization chamber for proton measurements (IBA PPC05) was placed at the isocenter for dose and spill length verification. The chamber was connected to the IBA DoseX electrometer equipped with an API measuring at a rate of 1 kHz. To analyze dose rate dependence, measurements were normalized twice by the dose measured via the PPC05 and the average of the specific measurement group. The integration time was kept the same at 10 s for all measurements. At higher dose rates and the same doses, the normalized charge increases until it reaches a plateau, then saturates in the flash regime. This is expected since, at higher current, charge traps get filled the higher the dose rate, until all the charge traps saturate. To test for radiation hardness, dose sensitivity was measured before and after heavy irradiation of 3 kGy using a clinical map. The sensitivity was reduced by 10%. The results were reproducible and linear with dose.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A single radiation hard pixel detector for proton dosimetry, imaging, energy resolution and beam monitoringarXiv · 2026
- Combining proton FLASH and spatially fractionated radiotherapy: experimental and simulation-based dosimetric characterizationPhysics in Medicine and Biology · 2026
- Comparative analysis of the responses of EBT-XD and EBT4 films to ultra-high dose rate proton beamsRadiation Physics and Chemistry · 2025
- X-ray FLASH: Technologies, evidence and challenges to translation.The British journal of radiology · 2026
- Dynamic Range Modulation via Integrated Universal Range Shifter and Ridge Filter: Enabling Simplified Delivery for FLASH-Enabled IMPT.Physics in medicine and biology · 2026
- Comparison of radiation damage effects in plastic scintillators after ultra-high dose rate irradiations of 9-MeV electrons, 200-MeV electrons, and 87-MeV protonsJournal of Physics: Conference Series · 2026