Design of a front-end signal conditioning circuit for space neutron detection payloads based on 3 He gas detectors
Shihao Liu, Zhi Hou, Qinggang Su, Tao Jing, Chongyi Wang
Key points
From the paper's conclusion- Addressing the engineering requirements for multi-channel integration in micro-nano satellite space detection instruments, this paper develops a front-end signal conditioning circuit tailored for 3 He gas detectors to facilitate space neutron energy spectrum analysis.
Abstract
The ongoing advancement of commercial aerospace technology has highlighted the diverse impacts of high-energy space radiation on commercial off-the-shelf (COTS) components, radiation detection systems, and space radiation dosimetry. In radiation detection, the signals output by detectors are inherently weak, necessitating front-end circuits with low noise and high gain for preliminary signal processing. Addressing the engineering requirements for multi-channel integration in micro-nano satellite space detection instruments, this paper develops a front-end signal conditioning circuit tailored for 3 He gas detectors to facilitate space neutron energy spectrum analysis. The proposed circuit integrates a charge-sensitive preamplifier, a pole-zero cancellation network, a filtering and shaping circuit, and a proportional amplifier. These components sequentially execute charge integration and pulse tail suppression, ultimately generating a quasi-Gaussian pulse. This provides a stable and reliable input for subsequent energy spectrum analysis processors. Circuit simulation tests and actual measurement results using an Am-Be neutron source demonstrate that the designed circuit meets the operational requirements for space nuclear radiation detection using a 3 He gas detector.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance of the front-end trigger detector for the High-energy Proton Beam Experimental Station (HPES) at CSNS-IIJournal of Instrumentation · 2026
- Optimisation of gas mixture for charge and light readout with Micromegas detectorsarXiv · 2026
- FPGA-Based Automatic Digital Pole-Zero Cancellation for Nuclear Pulse Spectroscopy Systems: Real-Time Decay Constant Estimation and Adaptive Tuning.Sensors (Basel, Switzerland) · 2026
- Development of a two-channel digital acquisition system for Δ E-E particle identificationJournal of Instrumentation · 2026
- Dual-Mode Multi-Microphone Enhanced Photoacoustic Gas Sensing System Using a Waveguide Ring ArrayAnalytical Chemistry · 2025
- On–off-beam differential resonance quartz-enhanced photoacoustic spectroscopy gas sensing technologyOptics Letters · 2026