Journal of Physics: Conference Series

Emissivity measurement of an N-type thermocouple

Yuhao Lv, Peiyong Wang

Key points

From the paper's conclusion
  • The results show that the emissivity of the N-type thermocouple decreases nonlinearly from 0.970 to 0.715 over the temperature range of 823.7–1084.6 K, corresponding to an overall reduction of approximately 26.3%.

Abstract

Radiative heat transfer from the thermocouple introduces errors in high-temperature gas temperature measurements, making thermocouple emissivity a key parameter for radiation correction. In this study, temperature measurements were performed using a 0.5 mm N-type thermocouple in lean Hencken flames. Combined with CFD simulations, the thermocouple emissivity was determined by iteratively adjusting its value until the simulated junction temperature agreed with the experimental measurements. The results show that the emissivity of the N-type thermocouple decreases nonlinearly from 0.970 to 0.715 over the temperature range of 823.7–1084.6 K, corresponding to an overall reduction of approximately 26.3%. These results demonstrate that the temperature dependence of thermocouple emissivity should be considered in the radiation correction for gas temperature measurements. The emissivity characteristics obtained in this study provide a useful reference for correcting the temperature measurement errors of N-type thermocouples under oxidizing conditions.

Read the paper