Systematic Underestimation of Rectangular Input Volumes Measured by Pressure Pipe Flowmeters
Luisfilippo Lanza, Umberto Sanfilippo
Key points
From the paper's conclusion- These two effects, together, lead to a systematic underestimation of measured flow volumes in comparison to their real values.
Abstract
The rectangular flow inputs generated by either a pump or a valve, even if with instantaneous beginning and end, are not measured as such by a flowmeter. In fact, there is always a mechanical (e.g. turbine flowmeters) or numerical (e.g. ultrasonic, electromagnetic and Coriolis flowmeters) damping effect. Moreover, in general, an instrument provides non-zero values only for flow rates above a minimum threshold. These two effects, together, lead to a systematic underestimation of measured flow volumes in comparison to their real values. Assessing these underestimations allows to improve the accuracy in calibration laboratories where a valve (or a pump) starts and stops flowmeter calibration tests with either volume or mass collection in downstream tank. This can also offer, to water companies, a method to estimate non-billed volumes due to intermittent demands, especially in case of low flow rates. This paper shows, analytically, how flowmeter’s readings respond to a singular rectangular flow input. In particular, the analytical discussion is developed for both instruments where the minimum flow rate threshold (cut-off) is determined by their intrinsic mechanical characteristics (e.g. turbine flowmeters) and instruments with a minimum cut-off set as a software parameter (e.g. ultrasonic, electromagnetic and Coriolis flowmeters). The main results obtained by means of the analytical discussion are then summarized in dimensionless form. It is also presented an experimental case study carried out for a Coriolis flowmeter. Finally, it is addressed the estimation of the combined uncertainty of unmeasured flow volumes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Field-relevant metrological performance of water meters: a topical review for water-infrastructure monitoringMeasurement Science and Technology · 2026
- A microfluidic vacuum-pressure-driven device for liquid viscosity measurementScientific Reports · 2026
- The Accuracy and Precision of Gravity-Driven Infusions.Prehospital emergency care · 2026
- Transit-time jitter and bandwidth-limited uncertainty in ultrasonic flow sensing under valve-induced turbulence in pump-controlled systems.Ultrasonics · 2026
- Volume-of-Fluid Simulations of Intermittent Gas–Liquid Flow: Physical Credibility Assessment and Benchmarking of Image-Based Velocity MeasurementModelling · 2026
- Design, Implementation and Functional Validation of Raspberry Pi 5–MATLAB Dual-Channel Telemetry System for Graywater Process-Flow MonitoringApplied Sciences · 2026