Journal of neural engineering

NeuroTest: a benchtop testbed for evaluating sensing-capable electrophysiology and neurostimulation systems.

Jeremiah K Morrow, Hanbin Cho, Moaad Benjaber, Timothy J Denison, Alik S Widge, Jeffrey A Herron

Psychiatry and Behavioral Sciences, University of Minnesota, 2001 6th St SE, Minneapolis, Minnesota, 55455, United States. · University of Washington, 1959 NE Pacific Street, Box 356470, Seattle, Washington, 98195, United States.

Computational

Key points

From the paper's conclusion
  • We demonstrate the use of the NTB to assess electrophysiology systems in two separate applications.

Abstract

The development of increasingly complex implantable neuromodulation devices requires extensive benchtop validation and testing prior to in vivo implementation. Our goal is to enable this in vitro testing by creating a benchtop system that is accessible, easy-to-use, and functions reliably across academic and industry environments, reducing discordant results that arise from discordant approaches. Furthermore, such a system paves the way for standardizing benchtop characterization of future implantable systems by establishing a repeatable and comparable methodology for validating these devices and their capabilities. Approach. We describe the NeuroTest Board (NTB), a low-cost, open-source device that combines a low-noise current source waveform generator with a high-precision data acquisition system. We provide information on its architecture, operation, and performance as an initial benchmark of the potential capabilities of the system. Main results. We demonstrate the use of the NTB to assess electrophysiology systems in two separate applications. In one case, we assess a commonly used laboratory system (the RHD2132 headstage and Open Ephys data acquisition system), while the second case characterizes an investigational neuromodulation device in development for human use (the CorTec Brain InterChange). Following this initial sensing characterization, we explore stimulation-related artifacts that would hamper in vivo data collection. With the NTB, we develop and evaluate artifact mitigation strategies for each system, thus demonstrating an example application of this benchtop testbed. Significance. The ability to test neuromodulation devices against a common benchmark will enable faster development of novel therapeutics by increasing inter-institutional reliability and decreasing troubleshooting time.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers