Electrical characterization and accelerated aging of amorphous silicon carbide implantable encapsulation
Christopher K. Nguyen, Negar Geramifard, Yupeng Wu, Madhav Bhatt, Alexandra Joshi‐Imre, Sandeep Negi, Stuart F. Cogan
The University of Texas at Dallas · Blackrock Microsystems (United States)
Key points
From the paper's conclusion- We aimed to evaluate thin-film a-SiC encapsulation using electrical-accelerated aging (EAA) and to demonstrate a methodology for obtaining acceleration factors for EAA by Weibull analysis.
Abstract
Abstract Objective . Chronically implanted microelectrode arrays (MEAs) are used for stimulating and recording neural activity in research and clinical settings. However, their reliability can be compromised by insufficient encapsulation stability. Amorphous silicon carbide (a-SiC), a chemically stable and biocompatible material, has emerged as a potential thin-film encapsulation for MEAs. We aimed to evaluate thin-film a-SiC encapsulation using electrical-accelerated aging (EAA) and to demonstrate a methodology for obtaining acceleration factors for EAA by Weibull analysis. Approach . Interdigitated electrodes (IDEs) encapsulated with a-SiC were subjected to voltage cycling and stepped-voltage protocols to measure leakage currents in buffered saline at 37 °C. EAA employed incrementally increasing voltage biases over time to induce degradation and reveal failure mechanisms. Main results . IDEs exhibited a significant change in electrical behavior on exposure to saline, with failure initiating at specific voltages and accompanied by gas evolution at defect sites. Incremental voltage biasing revealed a capacitive-to-faradaic transition in leakage current response that was used as a failure criterion. Significance . Acceleration factors for voltage-driven accelerated aging of a-SiC thin-film encapsulation can be obtained by Weibull analysis using a mechanistic failure criterion. Breakdown occurs at processing-related defects in the a-SiC. This study demonstrates the use of EAA for evaluating failure in a-SiC thin-film encapsulation used in implantable MEAs. EAA is broadly applicable to thin-film MEAs and provides a highly relevant method of predicting implanted lifetimes of bioelectronics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Long-term in vitro and in vivo biocompatibility evaluation of encapsulation materials for micro-LEDs using a multimodal assessment strategy.Biomedical engineering letters · 2026
- Flexible Implantable Silicon Carbide Multi‐Electrode Array (MEA) for Cell Multichannel Recording and Cell AblationAdvanced Materials Technologies · 2026
- Thermal-mechanical coupling simulation and experimental validation of high-temperature encapsulation for wear debris sensorJournal of Physics: Conference Series · 2026
- Fatigue Resistance of Cardiac Lead Insulations - A Novel in vitro Test Bench Mimicking Physiological LoadingCurrent Directions in Biomedical Engineering · 2026
- ALI MEA60: A porous thin-film microelectrode array for air-liquid interface electrophysiology of 3D excitable tissuesbioRxiv · 2026
- On the reliability of non-hermetically encapsulated ceramic printed circuit boards as bridging elements in neural implants.Frontiers in neuroscience · 2026