  • On this basis, an automated Soleil compensator-based LWIR phase retardation measurement system was constructed.

Abstract

Conventional phase retardation measurement methods, such as the rotating-waveplate technique and spectral scanning method, are insufficient to meet the high-precision measurement requirements in the long-wave infrared (LWIR) band (8–14 μm). To address this issue, we designed and developed a high-precision phase retardation measurement system for the LWIR band based on Soleil compensator method. By using cadmium selenide (CdSe) birefringent crystals, the key parameters of the compensator were determined through geometric optical design and precision analysis, and a LWIR Soleil compensator was successfully fabricated at first. On this basis, an automated Soleil compensator-based LWIR phase retardation measurement system was constructed. The system was used to perform actual measurements of CdSe waveplates under test over the wavelength range of 8–14 μm. The measurement results are in good agreement with the classical theoretical values of Bhar (1976), with an average relative error of 0.16%, verifying the high precision and reliability of the system. This work provides core components and technical support for the polarization calibration and high-precision magnetic field measurement of the AIMS telescope, as well as for other LWIR polarization applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers