Physica Scripta

Fabrication and characterization of ZnBO-based ultrasonic transducers for potential cardiovascular sensing applications

Gökhan Ekinci, Fatih M Coşkun, Sümeyye Veziroglu

Key points

From the paper's conclusion
  • The electrical characterization of the fabricated ZnBO-based ultrasonic transducers was performed autonomously using a LabVIEW interface under a 500 kHz sinusoidal excitation wave.

Abstract

Cardiovascular diseases, driven primarily by hypertension and atherosclerosis, remain a leading cause of mortality worldwide. Long-term continuous monitoring of blood pressure waveforms is essential for early diagnosis and clinical intervention. In this study, we report the first fabrication and characterization of boron-doped zinc oxide (ZnBO)-based ultrasonic transducers designed for invasive (e.g., catheter-based) and non-invasive (e.g., external sensors) cardiovascular blood pressure measurements. Transducer devices with a Si/Zn0.95-XB0.05O/Ti/Au stack architecture were fabricated on self-oxidizing Si(100) substrates via the sol-gel dip-coating method and magnetron sputtering. The surface morphology and elemental composition of the ZnBO films were examined in detail using scanning electron microscopy (SEM) and energy-dispersive X-ray (EDX) analysis. In addition, the crystal structure and phase composition of the ZnBO films were analyzed using X-ray diffraction (XRD). The electrical characterization of the fabricated ZnBO-based ultrasonic transducers was performed autonomously using a LabVIEW interface under a 500 kHz sinusoidal excitation wave. The fabricated biosensors generated substantial output voltages of up to +3 V and -4 V with minimal wave distortion, matching the transducer performance and -6 dB fractional bandwidth requirements demanded by advanced cardiovascular monitoring systems.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Fabrication and characterization of ZnBO-based ultrasonic transducers for potential cardiovascular sensing applications — Science Skim