Physiological Measurement

Fused electrical impedance tomography using an electromagnetically tracked handheld probe for oral cancer margin assessment: reconstruction methods and phantom validation

Sophie Ann Lloyd, Ehsan Nasiri, Allaire F Doussan, Ryan J Halter, Ethan K Murphy

Key points

From the paper's conclusion
  • Overall, this result is a step towards a clinically deployable impedance imaging approach to scanning the entire tumor boundary, which could significantly help to improve surgical outcomes for OSCC.

Abstract

Objective: Positive margins in oral squamous cell carcinoma (OSCC) resection are common and significantly impact patient outcomes. Electrical impedance tomography (EIT) can be used to intraoperatively distinguish cancer from healthy tissue. The objective of this study was to develop and identify a fused EIT approach to accurately reconstruct a tumor boundary using multiple impedance measurements recorded with a handheld electrical impedance probe. Approach: Simulations and noise analyses were used to investigate the best reconstruction method for fused EIT, and a phantom experiment was conducted to evaluate the EIT reconstructions. Main results: Fused approaches accurately recovered the conductivity boundary from multiple probe sites, reducing the experimental edge-localization error from 0.76 ± 0.51 mm for single-site reconstructions to 0.04–0.20 mm for the fused methods. All fused methods achieved AUCs ≥ 0.98 for separating the two conductivity regions in both simulated and experimental data, compared with 0.96 ± 0.04 for single-site experimental reconstructions. In noise analyses, fused diffEIT was the most robust to added Gaussian noise, with FWHM and 10–90% edge width changing by at most 0.06 mm with up to 100% added noise, while SR was the least affected by simulated EM localization error, with all edge metrics changing by at most 0.1 mm. Significance: The measured multi-site probe experiments represent important steps in the development of an approach that can combine EIT from multiple locations to increase boundary localization accuracy. Overall, this result is a step towards a clinically deployable impedance imaging approach to scanning the entire tumor boundary, which could significantly help to improve surgical outcomes for OSCC.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers