Directional discordance between a deployed commercial 12-lead ECG interpretation algorithm and expert cardiologists, and its distribution by sex and age: a post-deployment audit using PTB-XL.
Kaihui Du, Xiao Luo, Qibao Zhou
Department of Cardiovascular Medicine, Jiujang No 1 Peoples Hospital, jiujiang, Jiujiang, Jiangxi, 332000, China. · Jiujang No 1 Peoples Hospital, jiujiang, Jiujiang, 332000, China.
Key points
From the paper's conclusion- We characterized the directional discordance of the GE Marquette 12SL algorithm against expert labels.
Methods: In this retrospective audit of 14,885 ECGs from 13,480 patients (PTB-XL/PTB-XL+ dataset), we compared 12SL (v23) statements with standardized cardiologist labels for myocardial infarction (MI), conduction disturbance (CD), and left ventricular hypertrophy (LVH).
Abstract
Automated 12-lead ECG algorithms are widely used, yet their disagreement with cardiologists across patient subgroups is rarely audited on public data. We characterized the directional discordance of the GE Marquette 12SL algorithm against expert labels. Methods: In this retrospective audit of 14,885 ECGs from 13,480 patients (PTB-XL/PTB-XL+ dataset), we compared 12SL (v23) statements with standardized cardiologist labels for myocardial infarction (MI), conduction disturbance (CD), and left ventricular hypertrophy (LVH). We calculated observed agreement and paired marginal differences with 95% confidence intervals. Discordance was modeled using generalized estimating equations incorporating age, sex, and their interaction. Results: Observed agreement was high (0.842-0.896) but directionally divergent. 12SL produced fewer positive statements than cardiologists for CD (paired marginal difference 0.092, 95% CI 0.085-0.098) and more for LVH (-0.070, 95% CI -0.075 to -0.064); MI showed no net difference. The false-negative proportion for CD decreased with age, while false-positive proportions rose with age for CD and MI. For LVH false positives, the age association was strongly modified by sex (interaction 1.60, 95% CI 1.47-1.73), being absent in women but steep in men. The LVH direction was specific to primary 12SL statements and reversed if supportive statements were included. Conclusion: A widely deployed commercial ECG algorithm exhibits diagnosis-specific, directional discordance with expert reading that varies by age and sex, underscoring the value of reproducible post-deployment auditing. Because no clinical gold standard was available, these findings describe agreement with experts rather than absolute diagnostic accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- pedQTNet: A Deep Learning Approach to Estimate Corrected QT Intervals from Multi-Lead Conventional ECG Waveforms in Pediatric PatientsJournal of Medical Systems · 2026
- Diagnostic Performance of Automated versus Human ECG Interpretation for Prior Myocardial Infarction Using CMR as the Reference StandardEuropean Heart Journal - Digital Health · 2026
- Teaching an ECG model where to look does not make it more robust to missing leads: a pre-registered comparison of guideline-based attention guidance with random lead maskingmedRxiv · 2026
- Why blinding sex is not enough: equitable discrimination, unequal operation in machine-learning detection of myocardial infarction from the ECG.Journal of electrocardiology · 2026
- Artificial intelligence-based automated interpretation of images of electrocardiograms: development and multinational validation of ECG-GPTEuropean Heart Journal - Digital Health · 2026
- Multimodal large language models demonstrate consistently low accuracy for 12-lead electrocardiogram interpretation: a 599-case benchmark studyEuropean Heart Journal · 2026