Physics in Medicine & Biology

Spatiotemporal Gaussian representation-based dynamic reconstruction and motion estimation framework for time-resolved volumetric MR imaging (DREME-GSMR)

Jiacheng Xie, Hua-Chieh Shao, Can Wu, Ricardo Otazo, Jie Deng, Mu-Han Lin, Tsuicheng D Chiu, Jacob Buatti,

Key points

From the paper's conclusion
  • Representing patient anatomy and associated motion fields as 3D Gaussians, we developed a spatiotemporal Gaussian representation-based framework (DREME-GSMR), which enables time-resolved dynamic MRI reconstruction from a pre-treatment 3D MR scan without any prior anatomical/motion model.

Abstract

Objective. Time-resolved volumetric MR imaging that reconstructs a 3D MRI within sub-seconds to resolve deformable motion is essential for motion-adaptive radiotherapy. Representing patient anatomy and associated motion fields as 3D Gaussians, we developed a spatiotemporal Gaussian representation-based framework (DREME-GSMR), which enables time-resolved dynamic MRI reconstruction from a pre-treatment 3D MR scan without any prior anatomical/motion model. Approach. DREME-GSMR represents a reference MRI volume and a corresponding low-rank motion model (as motion-basis components) using 3D Gaussians, and incorporates a dual-path MLP/CNN motion encoder to estimate temporal motion coefficients of the motion model from raw k-space-derived signals. Using the solved motion model, DREME-GSMR can infer motion coefficients directly from new online k-space data, enabling subsequent intra-treatment volumetric MR imaging and motion tracking (real-time imaging). A motion-augmentation strategy is further introduced to improve robustness to unseen motion patterns during real-time imaging. Main results. DREME-GSMR was evaluated on the XCAT digital phantom, a physical motion phantom, and MR-LINAC datasets acquired from 6 healthy volunteers and 20 patients (with independent sequential scans for cross-evaluation). DREME-GSMR produced time-resolved MRIs at a ~400 ms temporal resolution, with an inference time of ~10 ms/volume. In XCAT experiments, DREME-GSMR achieved mean(s.d.) SSIM, tumor center-of-mass error (COME), and Dice similarity coefficient (DSC) of 0.92(0.01)/0.91(0.02), 0.50(0.15)/0.65(0.19) mm, and 0.92(0.02)/0.92(0.03) for dynamic reconstruction/real-time imaging. For the physical phantom, the mean target COME was 1.20(0.94)/1.40(1.15) mm for dynamic/real-time imaging, while for volunteers and patients, where no ground truth is available, the mean real-time liver COME relative to a dynamic reconstruction-derived pseudo-ground truth was 1.31(0.82) and 0.96(0.64) mm, respectively. Significance. Evaluated on digital simulations and MR-LINAC measurements, DREME-GSMR enables sub-second 3D dynamic MRI reconstruction and real-time volumetric motion tracking from stack-of-stars k-space data, supporting intra-treatment motion-adaptive MR-guided radiotherapy.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.