First multi-institutional systematic comparison of the neutron ambient dose equivalent produced by proton therapy systems
Arash Darafsheh, Yao Hao, A. Bey, Michele M. Kim, Eric S Diffenderfer, Lingshu Yin, Chin-Cheng Chen, Zhiyan Xiao,
Washington University in St. Louis · University of Pennsylvania
Key points
From the paper's conclusion- We present the first comprehensive comparison of neutron ambient dose equivalent (H*(10)) produced by clinically operational proton therapy systems.
Abstract
Abstract Objective. Isochronous cyclotrons, synchrocyclotrons, and synchrotrons are used to accelerate protons for proton therapy. An accurate measurement of neutron doses generated by these accelerators and associated delivery systems and its clinical relevance requires systematic protocols and proper neutron dosimetry for a meaningful assessment. We present the first comprehensive comparison of neutron ambient dose equivalent (H*(10)) produced by clinically operational proton therapy systems. Approach. Treatment plans with 10 cm modulation-depth and ranges of 10 cm (R10M10) and 25 cm (R25M10) were created to cover a 10 × 10 × 10 cm 3 water target. The pencil beam scanning proton therapy machines studied were: two gantry-mounted synchrocyclotrons (Hyperscan, Mevion, half-gantry), two isochronous cyclotrons (ProBeam, Varian, full-gantry), one isochronous cyclotron (Proteus, IBA, full-gantry), and two synchrotrons (PROBEAT, Hitachi, full- and half-gantry). Proton beams were delivered to 30 × 30 × 40 cm 3 plastic water phantoms. WENDI-II and LUPIN-BF3-NP neutron rem-meters were positioned at three angles (0°, 45°, 90°) relative to the beam direction to measure the neutron H*(10) at distances between 50–300 cm from the isocenter. Main results. H*(10) showed dependence on beam energy, machine type, and measurement location. The highest reading was for the gantry-mounted synchrocyclotron, whereas other systems produced approximately comparable neutron doses. In all cases, the H*(10) reduced with distance from the isocenter. The H*(10) drop at 2 m distance compared to that at 0.5 m was a factor of ∼5 for the gantry-mounted synchrocyclotron whereas in other systems the decrease was a factor of 10. The WENDI-II device suffered from dead-time-associated under-estimation of the dose by a factor of ∼2–3 under the synchrocyclotron beam due to its high dose-per-pulse. However, WENDI-II and LUPIN-BF3-NP results were within reasonable agreement in isochronous cyclotron and synchrotron beams, indicating that both devices are suitable for those systems. Significance. Neutron H*(10) is dependent on various parameters including beam energy, measurement location, as well as machine design. Caution must be exercised in choosing the appropriate neutron-dose-measurement device to be used for low-duty-factor, particularly in high-instantaneous-rate proton delivery systems. By delivering the same volumetric proton dose across different machines, this work provides a benchmark for inter-system comparisons and serves as a foundation for future studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Practical considerations in measuring out-of-field neutron dose for pulsed and gantry-mounted proton therapy machines.Physics in medicine and biology · 2026
- Angular Distributions of Secondary Neutron and Photon Ambient Dose Equivalent in Proton Therapy: A Systematic PHITS 3.35 Study at 70-250 MeV.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Evaluation of neutron dose equivalent in prostate stereotactic body radiotherapy by bubble detector.Biomedical physics & engineering express · 2026
- Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT fieldJapanese Journal of Applied Physics · 2026
- Towards real-time ion range verification via hybrid Compton-PET 3D imaging at isochronous cyclotronsarXiv · 2026
- Radiation-induced tumor following pencil beam scanning proton therapy for pediatric intracranial ependymoma: a pragmatic analysis of risk factors.Journal of radiation research · 2026