Signal decomposition methods for non-contact vital-sign extraction using stepped-frequency continuous-wave radar
Piyush K Mathurkar, Abhay N Gaikwad
Key points
From the paper's conclusion- Validation of the signal processing pipeline was done using a 1.5 Hz mechanical oscillating machine, which yielded a 1.33% relative error; a measurement uncertainty analysis (coverage factor k = 2) established the reported rate as 18.8 ± 0.7 breaths min –1 and 66.8 ± 1.2 beats min −1 for Subject 1, and 14.4 ± 0.7 breaths min –1 and 76.5 ± 1.2 beats min −1 for Subject 2.
Abstract
Non-contact methods that accurately estimate heart and respiration rates are an important goal in radar-based vital sign measurement that can be used for health surveillance without intrusion, sleep monitoring, and medical scenarios where contact sensors are not feasible. This paper conducts a comparative study of three signal decomposition techniques—empirical mode decomposition (EMD), ensemble empirical mode decomposition (EEMD), and variational mode decomposition (VMD)—for simultaneous estimation of cardiac and respiratory rates from stepped-frequency continuous-wave radar signals. The radar echo was processed by coherently averaging the range bins covering the subject’s chest to form respiratory and cardiac displacement signals that were decomposed into intrinsic mode functions. The dominant physiological mode in each target band (0.10–0.50 Hz for respiration, 0.80–2.00 Hz for the cardiac component) was selected by the in-band signal-to-noise ratio (SNR) with spectral entropy as the descriptor of spectral compactness. The signal processing pipeline was evaluated on two subjects at rest. The three methods agreed closely, with the respiratory rate and heart rate for Subject 1 being 18.8 breaths min –1 and 66.8 beats min −1 , respectively, and 14.4 breaths min –1 and 76.5 beats min −1 for Subject 2. VMD achieved the highest in-band SNR in both channels (23.89 dB for respiratory, 12.21 dB for cardiac), exceeding EMD and EEMD by 7 dB and 4.5 dB, with the lowest spectral entropy (0.215 and 0.457, a 29%–38% reduction). VMD also required only four modes versus five or six for the others, simplifying selection and improving reproducibility. Validation of the signal processing pipeline was done using a 1.5 Hz mechanical oscillating machine, which yielded a 1.33% relative error; a measurement uncertainty analysis (coverage factor k = 2) established the reported rate as 18.8 ± 0.7 breaths min –1 and 66.8 ± 1.2 beats min −1 for Subject 1, and 14.4 ± 0.7 breaths min –1 and 76.5 ± 1.2 beats min −1 for Subject 2.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Radar-Based Heart Rate Estimation Method Under Respiratory Harmonic Interference.Sensors (Basel, Switzerland) · 2026
- Respiration and Heartbeat Detection Using FMCW Radar Based on Secondary Variational Mode DecompositionEng · 2026
- Millimeter-wave radar for non-contact heart rate monitoring: Composite filtering and multidimensional evaluation.Technology and health care : official journal of the European Society for Engineering and Medicine · 2026
- A Synchronized Tri-Modality Wireless System for Multi-Point Non-Contact Cardiac MonitoringarXiv · 2026
- Embedded Cardiorespiratory Monitoring Using a 77 GHz FMCW Radar: On-Chip Multi-Timescale Tiny 1D CNN and Whole-Node Power Evaluation.Sensors (Basel, Switzerland) · 2026
- DuBCG: Unlocking In-Home, Clinically Comparable mmWave Cardiac Monitoring for Co-Sleepers via Physics-Informed Spatial SensingProceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies · 2026