Mechanically-induced fiducial marker system for event-based synchronization of independent recording instruments.
Kourosh Kalayeh, John O L DeLancey, James A Ashton-Miller, J Brian Fowlkes
Department of Radiology, University of Michigan, Ann Arbor, MI, United States of America. · Department of Urology, University of Michigan, Ann Arbor, MI, United States of America.
Key points
From the paper's conclusion- The approach generalizes to instrument pairings in which one instrument provides a stable electrical trigger with a known relation to its recording timeline and the other is a mechanically-sensitive receiver of adequate bandwidth, provided the marker does not contaminate the signal of interest.
Abstract
Many experimental and clinical measurements require simultaneous acquisition from independent instruments that do not share a clock. We describe a device that enables such pairings via event-based synchronization: an electrical trigger from one instrument drives an Arduino-controlled stepper actuator that produces a brief mechanical event at a location sensed by the other instrument, leaving a fiducial marker for post-hoc temporal alignment. The motivating application is synchronization of high-resolution manometry (HRM) with transperineal ultrasound for phenotyping stress urinary incontinence. Bench characterization yielded a device latency of 802.5 μ m ± 46.2 μ m (mean ± SD, n = 30 ), of which the electromechanical response of the actuator accounts for ≈ 84 % of the mean latency and the dominant share of the trial-to-trial variation (i.e. the SD of 46.2 μ m ). For the ultrasound-HRM application, the synchronization uncertainty is ≈ 6 ms ( n = 30 markers), set by the HRM rather than the device, whose intrinsic jitter contributes negligibly. The approach generalizes to instrument pairings in which one instrument provides a stable electrical trigger with a known relation to its recording timeline and the other is a mechanically-sensitive receiver of adequate bandwidth, provided the marker does not contaminate the signal of interest.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Synchronizing EEG with Intracranial DBS Electrode Recordings for Neurophysiological ResearchmedRxiv · 2026
- Validation of video-based event detection against laboratory-standard biomechanical measures: a synchronized multimodal study.Frontiers in bioengineering and biotechnology · 2026
- DBSsync: combining intracranial and multimodal data to investigate new biomarkers in Parkinson’s diseasenpj Parkinson s Disease · 2026
- Quantifying Soft‐Surgery in Cochlear Implantation: Multimodal Data From 30 International SpecialistsThe Laryngoscope · 2026
- High-Precision Insertion Force Measurements for Cochlear Implant Surgery: Accuracy Analysis of an Improved DeviceCurrent Directions in Biomedical Engineering · 2026
- Objective motion-based performance assessment in STA-MCA microanastomosis: a postprocedural debriefing platform.Neurosurgical focus · 2026