Query-based prediction of 3D x-ray scatter fields in an interventional suite using a transformer model.
Niklas Rettig, Patrick Schülein, Yannick Bukschat, Marcus Vetter
Institute for Applied Artificial Intelligence and Robotics (A2IR), Technical University of Applied Sciences Mannheim, Mannheim, Germany.
Key points
From the paper's conclusion- We present a transformer-based model for scatter dosimetry in a single simulated interventional suite with fixed room geometry and object inventory, supporting near real-time inference on localised subvolumes and interactive runtimes for whole-room evaluation.
Abstract
Fluoroscopy-guided interventions expose clinical staff to x-ray scatter radiation, motivating accurate three-dimensional scatter-field predictions for radiation protection planning. Monte Carlo (MC) simulations provide reference accuracy but require minutes to hours of computation, while analytical models are faster but often lack spatial fidelity in complex geometries. We present a transformer-based model for scatter dosimetry in a single simulated interventional suite with fixed room geometry and object inventory, supporting near real-time inference on localised subvolumes and interactive runtimes for whole-room evaluation. We train a separate transformer-based model for each target quantity: air kerma per source photon, transported energy, and energy fluence per source photon. Each predictor encodes spectrum and technique parameters and represents room objects and their poses as a set of scene tokens. Predictions are obtained by coordinate-conditioned regression at queried voxel locations, so inference can be restricted to selected subvolumes and runtime scales with the number of queried voxels. The training set comprisessimulated configurations with varied C-arm poses, spectra, patient-table and patient phantom positions. On an independent test set ofconfigurations, the model achieves medianfor log-transformed targets and structural similarity around. In linear units, medianvalues arefor air kerma per source photon,for transported energy, andfor energy fluence. Full-room inference, reconstructing the effective grid of aboutmillion voxels, takeson an NVIDIA L40S GPU compared tofor MC simulation, yieldingaspeed-up. These results represent a step toward interactive, room-scale radiation protection planning, enabling rapid what-if analyses of staff positioning and procedure planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fast operating room scattered radiation calculation in x-ray guided interventions by using deep learningJournal of Radiological Protection · 2025
- Physics-informed machine learning-based dose calculation for patients undergoing fluoroscopically guided interventional procedures.Physics in medicine and biology · 2026
- Latent-aligned scalable interpolant transformer for controllable chest X-ray synthesisScientific Reports · 2026
- Differentiable surrogate for detector simulation and design with diffusion modelsMachine Learning Science and Technology · 2026
- Enhanced diffusion model with 3D structural and segmentation priors for lung cancer radiotherapy dose prediction.Physics in medicine and biology · 2026
- Grounding Radiology Report Findings into Medical Image Segmentation.NPJ digital medicine · 2026